If the preceding tips do not
yield the desired results, or
for any other defects or prob-
lems not mentioned here, please pull
the power plug and contact our Ser-
vice Center immediately.
Do not open the Hot Air Gen-
erator! Without special tools
and specialized knowledge,
any attempt to open the device will
cause damage to it. Such an attempt
will invalidate the warranty claim.
Do not try to remedy the defect your-
self. Contact directly our Service
Center.
10.7. Technical Specifications
VOLCANO MEDIC 2

Vaporization System

Voltage:
220-240 V / 50-60 Hz (alternatively
available with 110-120 V / 50-60 Hz)
The voltage specification is located
on the bottom side of the
VOLCANO MEDIC 2 Hot Air Generator.
Power consumption: 270 VA
Air pump output: 10 VA
Vaporization temperature: adjustable
between approx. 40 - 210°C (104 -
410°F)
Air flow: approx. 12 l/min
Valve Balloon Size:
approx. 12.5 liters
10. H O T A I R G E N E R AT O R
123
Size of the Hot Air Generator:
20.0 x 18.0 cm (7.9 x 7.1 inch)
Weight of the Hot Air Generator with
Power Cord, Filling Chamber, and
Valve Balloon with Mouthpiece:
approx. 1.8 kg (3.9 lbs)
Safety category II
Automatic switch off after 30 minutes
Please unplug to disconnect from the
mains current.
Internal fuse F1: F-T 4 A (applies to
110-120 V and 220-240 V devices)
This fuse may be replaced only by our
Service Center.
Subject to technical changes.
Patents and designs:
www.vapormed.com/patents
Manufacturer:
Storz & Bickel GmbH
In Grubenäcker 5-9,
78532 Tuttlingen, Germany