7.5.2

Rinse: Day

May 2003
This procedure is used to rinse the wash solution system if the instrument is to be
left standing for a longer time. Alternatively it can be used to rinse the wash
solution system after the instrument has been left for a long time.
With this procedure you can use either distilled water or wash solution.
If the instrument is left to stand for 1 hr to 1/2 day, wash solution can be used, for
longer times use distilled water.
Rinse:
Day
-
+ other yes
Day:
CH1
-
+
exit yes
Rinse Solution ?
exit
yes
Waste Bottle OK?
exit
yes
Rinse ChX YY sec
Stop
Soaking
Stop
Run
XX:yyyyyyy
-
+ other yes
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
7. Maintenance and Cleaning
Select and confirm Day Rinse procedure
Select required channel (1 - 4)
Confirm rinsing solution and tube in bottle
Confirm waste bottle is empty
Wash system rinsed for set rinse time
The waste sink is filled with solution and
the manifold is lowered into it
Manifold left in solution until
Stop key pressed
Standby mode
Washing Procedure can be performed
7-11