Page 1
Instructions for Use 4K UHD 58" Monitor Important Please read the safety information and all information delivered with the product carefully to familiarize yourself with safe and effective usage.
Use of EIZO products WARNING EIZO products may only be used for the applications described in the catalog and in the relevant tech- nical documentation. If products and components from other manufacturers are used, these must be recommended or approved by EIZO. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
ROM provided and can be downloaded from www.eizo-or.com. 1.2 Intended use The CuratOR EX5841 is intended for the display of still images and moving images in the medical environment, for example, during endoscopic examinations. The monitor is not suit- able for mammography.
2 Safety information 2.1 General safety instructions Correct and safe operation of EIZO devices assume professional transport, storage, installa- tion, and connection, as well as careful operation and service. The devices may only be used for applications for which they are intended.
Page 8
Only use power cables approved for China. These power cables are identified by the labels "CCC" or "CQC". Observe the country-specific regulations Observe all regulations of the country in which the device is used. CuratOR EX5841 Instructions for Use, 04/2020...
Page 9
If the device is brought into a room with a higher or rising temperature, condensed water will form in and on the device. Do not switch on the device until the condensed water has evaporated. Other- wise, the device could be damaged. CuratOR EX5841 Instructions for Use, 04/2020...
Page 10
Note that monitors can fail as a result of aging, and that image properties such as brightness, con- trast,and color value can change. Do not touch the monitor screen Due to mechanical pressure or electrostatic discharges, touching the screen can result in brief distur- bances to the image. CuratOR EX5841 Instructions for Use, 04/2020...
Note Image quality To maintain constant image quality, EIZO recommends cleaning the monitor on a regular basis and checking image properties in accordance with all applicable local regulations. CuratOR EX5841 Instructions for Use, 04/2020...
Note Keep the packaging material for subsequent transport of the device. Device The CuratOR EX5841 is a 8 MP 58" LCD Monitor for mounting in a ceiling suspension unit or wall mount. Product Order number...
Page 13
User-specific operating configurations can be saved in the monitor as user presets. Typical operating configurations are saved ex-factory as factory presets. The presents can be se- lected during operation using the OSD or the Preset key. CuratOR EX5841 Instructions for Use, 04/2020...
In the worst case, deposits are possible which become evident as dark spots in a white picture and result in deterioration of the luminance. • Protect the monitor from dust, for example through appropriate construction measures at the installation site. CuratOR EX5841 Instructions for Use, 04/2020...
Page 15
If the monitor is installed such that it can move, make sure that persons or objects in the facility are not endangered by the monitor's range of movement. Note During transport, use the original packaging or service packaging. CuratOR EX5841 Instructions for Use, 04/2020...
3. Remove the enclosed components, such as cables, manuals, and CD from the upper foam insert. 4. Remove the upper foam insert. 5. Lift the monitor out of the lower part of the carton. Hold it on the side and underside. CuratOR EX5841 Instructions for Use, 04/2020...
If the monitor is to be mounted in a holder other than a VESA 400x400, remove the prein- stalled screws and cover the screw holes with the adhesive foil included in the scope of delivery. Otherwise, protection class IP32 cannot be maintained. CuratOR EX5841 Instructions for Use, 04/2020...
Page 18
The maximum torque for attaching to the holder is 18 Nm. • The screws used for attaching to the holder must meet the following requirements: Thread Strength 8.8 in accordance with ISO 898-1 Insertion depth 16 mm ~ 20 mm CuratOR EX5841 Instructions for Use, 04/2020...
Device settings may only be adjusted by service personnel. NOTICE Disconnecting from line power Always set the power switch to "Off" before disconnecting the device from power. Other- wise the device could be destroyed. CuratOR EX5841 Instructions for Use, 04/2020...
The equipotential bonding connection on the device can minimize contact voltages from dif- ferent sources, for example, when the device is used in a system with other devices. CuratOR EX5841 Instructions for Use, 04/2020...
Grounding screw Fig.: Rear view without cover 5.2.2 Connection panel CAUTION Opening the connector panel cover Only service may open the connector panel cover. Patients must not be present when the cover is open. CuratOR EX5841 Instructions for Use, 04/2020...
Page 22
2 USB Type A (downstream) for connecting external USB devices. USB Type B 1 USB type B (upstream) for connecting a USB host (such as a PC). 1 jack plug (3.5 mm) without function CuratOR EX5841 Instructions for Use, 04/2020...
Do not use any other fuse. NOTICE Line voltage and frequency Before connecting the device, make sure the line voltage and frequency correspond to the specifications on the name plate. CuratOR EX5841 Instructions for Use, 04/2020...
3. If external devices need to be supplied with 5 V voltage and up to 1 A, connect them to the Hirose connectors. 4. Connect the power cable to the power connector. 5. Attach the cover over the connection panel. ð The monitor can now be put into operation. CuratOR EX5841 Instructions for Use, 04/2020...
Reduce the selected entry. • Scroll up in the menu. • Enlarge the selected entry. ENTER • Opens the next menu. • Jumps to the element to the right. • Performs the selected function. CuratOR EX5841 Instructions for Use, 04/2020...
If the operation LED does not light green after the equipment has been switched on and a video signal has been applied: • check the system for basic connection and operating errors before contacting service personnel. CuratOR EX5841 Instructions for Use, 04/2020...
7.2 Prohibited cleaning agents and disinfectants After extended use, the cleaning agents and disinfectants listed can lighten the paint or damage the polarizer. Agent class Tested cleaning agents and Further examples disinfectants Organic acids Bio-AntiBact med CuratOR EX5841 Instructions for Use, 04/2020...
Check the ventilation and improve trol is reduced by half. Brightness is re- conditions if necessary. duced significantly to lower the tempera- • Select a lower brightness level for ture and avoid potential damage. standard operation. CuratOR EX5841 Instructions for Use, 04/2020...
Internal error • Inform service department Dark LED timeout activated • No error. • Switch off the LED timeout setting. Operation LED is defective • Inform Service. CuratOR EX5841 Instructions for Use, 04/2020...
Operation Failed Execution of the requested oper- ation was canceled. Please close OSD first A USB stick with update software Service: Remove the USB stick. was connected with the OSD Close OSD. open. CuratOR EX5841 Instructions for Use, 04/2020...
< = 0.9 A @ 240 V / <= 2.0 A @ 100 V Max. power consumption 200 W/ 216 VA Energy Saving Mode 0 W (power switch off); < 5 W (voltage outputs off); <50 W (voltage outputs on) CuratOR EX5841 Instructions for Use, 04/2020...
IP45 (front); IP32 (back) according to EN 60529 Connection panel On rear panel, under cover Weight 45 kg +/- 5% Dimensions (W x H x D) in mm 1329 x 781 x 89 (without stand) CuratOR EX5841 Instructions for Use, 04/2020...
IEC 60601-1 • IEC 62368 • CAN/CSA-C22.2 No. 60601-1:14 • ANSI/AAMI ES60601-1 Protection class Protection class I Degree of protection IP45 (front); IP32 (back) Medical device classification Class I according to 93/42/EEC, Appendix IX CuratOR EX5841 Instructions for Use, 04/2020...
Dimension drawings 10.1 View from front and side 10 Dimension drawings All dimensions in mm 10.1 View from front and side 10.2 Rear view CuratOR EX5841 Instructions for Use, 04/2020...
Symbol for degree of protection according to DIN EN 60529. "On" symbol (voltage). "Off" symbol (voltage). Input for service calls. Symbol for “Comply with the instructions for use”. Symbol for "Authorized service personnel only". Symbol for "Hazardous electrical voltage". CuratOR EX5841 Instructions for Use, 04/2020...
IEC/EN 60601-1-2. Electromagnetic radiation The CuratOR EX5841 is intended for use in the electromagnetic environments noted below. Customers and users of the CuratOR EX5841 have to ensure that the device is used in such an en- vironment. Radiation test Conformity...
Page 38
The CuratOR EX5841 was tested with the following compliance levels in accordance with the test re- quirements for professional healthcare facilities, as established in IEC/EN 6061-1-2. Customers and users of the CuratOR EX5841 have to ensure that the monitor is used in such an en- vironment.
Page 39
11.2 Information on electromagnetic compatibility (EMC) Electromagnetic interference immunity The CuratOR EX5841 was tested with the following compliance levels in accordance with the test re- quirements for professional healthcare facilities, as established in IEC/EN 6061-1-2. Customers and users of the monitor have to ensure that the monitor is used in such an environment.
Page 40
Recommended minimum distance between portable or mobile RF communications devices and the CuratOR EX5841 The CuratOR EX5841 is intended for use in an electromagnetic environment in which interference due to electromagnetic radiation is controlled. For other portable and mobile RF communication de- vices (transmitters), the recommended minimum distance between the portable and mobile RF com- munication devices (transmitters) and the monitor applies as listed below.
Page 41
Recommended minimum distance between portable or mobile RF communications devices and the CuratOR EX5841 The CuratOR EX5841 is intended for use in an electromagnetic environment in which interference due to electromagnetic radiation is controlled. The customer or user of the monitor can help prevent electromagnetic interference by maintaining the recommended minimum distance between portable and mobile RF communications devices (transmitters) and the monitor.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note Use the attached specified cable below or EIZO signal cable with this monitor so as to keep interfer- ence within the limits of a Class B digital device. •...
EFUP shall be done in accordance with the Product Maintenance Procedures. This product must not be disposed of as unsorted municipal waste, and must be col- lected separately and handled properly after decommissioning. CuratOR EX5841 Instructions for Use, 04/2020...
Page 44
Applications of hazardous substances in this medical device are required to achieve its intended clinical uses, and/or to provide better protection to human beings and/or to environment, due to lack of reasonably (economically or technically) available substitutes. 产品中有毒有害物质或元素的名称及含量 Table of hazardous substances’ name and con- centration. CuratOR EX5841 Instructions for Use, 04/2020...
11.6 Declaration of compliance with India RoHS 11.6 Declaration of compliance with India RoHS We, EIZO Corporation, hereby declare and guarantee that this product has been designed and manufactured in compliance with the E-Waste management rule 2016 which prohibit the inclusion of the following substances except for the exemptions listed in schedule II.
11.10 Trademarks 11.10 Trademarks The EIZO Logo is a registered trademark of EIZO Corporation in Japan and other countries. EIZO is a registered trademark of EIZO Corporation in Japan and other countries. RadiForce is a registered trademark of EIZO Corporation in Japan and other countries.
12 General safety instructions 7 Grounding 19 Grounding screw 20 Installation site 14 Line power 23 Overheating 14 Power connector 23 Scope of delivery 12 Service personnel 6 Shielding measures 19 Switch on 26 CuratOR EX5841 Instructions for Use, 04/2020...
Need help?
Do you have a question about the CuratOR EX5841 and is the answer not in the manual?
Questions and answers