Download Log File - BreatheSafe INPRESS TS User Manual

Table of Contents

Advertisement

Controlled document: M-0002B
I N P R E S S T S
Revision: B
Issue Date: 30 August 2021
C on te nt s
1 USER GUIDE
1 Key Features
2 WARNING
3 CONTROLLER APPEARANCE
4 FUNCTIONS
5 OPERATING INSTRUCTIONS
5 Enter Setup Mode
6 Change Date and Time
6 Change Pressure Setpoint
6 Change ALarm Setpoint
7 Enable / Disable Buzzer
7 Change Alarm DELAY Settings
7 Change Service Reminders
8 Calibrate Zero Sensor
8 Reset Data Log
8 Test Inputs Option
9 TECHNICAL DETAILS
9 Map of Functions
11 General Arrangement
12 Connector Pinout Details
14 Installation
15 Fan Speed Control Output
15 Mounting
15 Supply Power
15 Calibration
15 Control Override
16 Data Logging Information
20 BREATHESAFE WARRANTY
21 CONTACT US
© BREATHESAFE 2021
DATA LOGGING INFORMATION
SET INTERVAL MINUTES
Use ARROW buttons to adjust the frequency of Data recordings.
CLEAR DATA LOG
Pressing the CLEAR DATA LOG button takes you to the PIN Entry
Screen (Factory Access PIN is required). Once the PIN is entered the
Unit clears and deletes the stored data.
The screen will show a progress bar during this step, then will return
to the "Factory Settings Screen" on completion.

DOWNLOAD LOG FILE

Pressing this button begins the download of the stored data to the
terminal software installed on your computer. The logging intervals
for the downloaded data will be based on the Set Interval.
Controlled document:
The screen will show a progress bar during this step, then will return
Issue Date:
to the "Factory Settings Screen" on completion.
GET REAL TIME LOG
Pressing this button begins the output of the log data to the terminal
software installed on your computer. The logging intervals for the real
time data will be every second. The button will change text to "STOP
REAL TIME LOG", and can be pressed to stop the feed.
SET UP FOR DATA RETREVAL
Using the serial connector on the wiring harness (or by fitting the
PAGE 16
www.breathe-safe.com.au
bridging harness with serial connector), connect your computer with
a serial cable.
Using terminal software similar to the freeware program TeraTerm
(easily found through an internet search).
In the software set up the connection as a SERIAL Connection via one
of your computer's COM Ports, ensuring your Serial settings are as
follows:
Baud Rate: 57600
Data: 8 bit
Parity: EVEN
Stop: 2 bit
Flow Control: NONE
DATA LOGGING SCHEDULE
The BreatheSafe INPRESS TS controller is designed to start data
logging as soon as it is powered up. Data samples are taken and
recorded at intervals with time, date and current pressure readings.
This is additional to normal alarm logging which occurs at every alarm
instance.
The unit is designed to log over 10900 instances in its internal
memory. The memory is not battery dependent and can retain data
for up to 10 years. Under normal operation the controller logs one
sample every hour.
This time period is shortened if the BreatheSafe INPRESS TS unit
experiences many alarms throughout each day.
When the data logger's memory is full it returns to the beginning
and over-writes previously recorded data in an endless loop. It is
therefore very important that the data is downloaded as required to
ensure that there is no loss of important information.