Card Level Measurements - 60G&80G-L Interface Boards; Checking Voltages; Checking Basic Functions; Checking Printing - Innomed Medical Heart Screen 60G Service Manual

Simultaneous 12-channel ecg devices with 58 / 80mm wide 1- / 3-channel built-in printer and lcd display
Hide thumbs Also See for Heart Screen 60G:
Table of Contents

Advertisement

Innomed Medical
7 Card level measurements –
60G&80G-L Interface Boards
7.1

Checking voltages

Connect the power supply set to 9V/1A to the connection of the accu pack on the Main Boar and connect the Interface
Board. Don't connect the printer to the Interface Board.
Check the switch on/off possibilities. Switch on and off the device by the on/off button.
Check the following signals:
Measurement point
J1.3
J1.2
GND can be found at the point of J1.4.
7.2

Checking basic functions

Check the following:
Test procedure
Backlight test
Contrast test
Keyboard test
7.3

Checking printing

7.3.1 With printer partially connected

1.
Connect a 4700uF capacitor in parallel with the power supply (9V max. current limit) substituting the battery.
2.
Connect only the printer motor connection and load the paper. Connect a short on the ECG input.
Test procedure
Burn impulse test
Simultaneously with the burn impulse test check the following:
Measurement place
Interface Board J1.2
3.
Stop supplying burn impulses. After approx. 3 seconds measure the burn voltage.
Measurement place
Interface Board J1.2

7.3.2 Print test with printer

Connect the burning head as well and start 50mm/sec on-line printing. Watch the signal shape, that the displayed
curves are continuous, there are no breaks, spikes, faded sections.
Further, measure the power consumption at the battery connection's negative point:
Measurement place
Main Board J4.2
2014.01.09
Heart Screen 60G-80G-L
Nominal value
Tolerance
2.5V
+/-0,1V
0V
+/-0,1V
Comment
Backlight is operational when the device is on.
The LCD contrast is acceptable, not foggy, the LCD is operational.
In the test menu (HS80G-L) or make sure the appropriate function is
performed when the buttons are pressed (HS60G)
Comment
Start from the test menu (HS80G-L) or with the REC button (HS60G).
Max. 5ms burn time, measure at J1.1 on the Interface Board (active signal at high
level).
Nominal value
Tolerance
7.2V
+/-0.1V
Nominal value
Tolerance
7.2V
+/-0.1V
Nominal value
1-2.5 A fluctuating
Comment
Vref
HEADPOWER
Comment
HEADPOWER
Comment
HEADPOWER
Comment
Current from the battery.
31/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
Service manual