8.3.6 Checking frequency transfer
Apply +DC signal onto the amplifier's input (2000 +/-0.5% RVpp / 150 Hz sine) from a ground independent generator
according to the table. DC must be +320, -320 mV.
The filter must be off.
Connect the patient cable according to the following tables and with the channel selector start the channels one-by-one,
and measure the signal amplitude or during 50 mm/sec *1 printing.
a/ In case of DC = -320 mV
The signal's amplitude on the examined lead must be min. 1400 RV.
P1
L
R
V1
V2
V3
V4
V5
V6
b. In case of DC = -320 mV
The signal's amplitude on the examined lead must be min. 1400 RV.
P1
L
R
V1
V2
V3
V4
V5
V6
8.3.7 Checking filter
a. "Line" filter on at 50Hz
Apply 4 mVpp 50Hz sine signal from the generator between the device's V2 and all other connected inputs. The filter
is acceptable if the dampening is less than 20dB when the frequency is changed in the 48.5 Hz - 51.5 Hz range. The
measurement should be performed on the V2 lead by making a recording at 50 mm/sec speed and normal amplification
with the filter on. The recorded signal must be less than 4 mm.
b. "Line" filter on at 60 Hz
Apply 4 mVpp 60Hz sine signal from the generator between the device's V2 and all other connected inputs. The filter
is acceptable if the dampening is greater than 20dB when the frequency is changed in the 59 Hz - 61 Hz range. The
measurement should be performed on the V2 lead by making a recording at 50 mm/sec speed and normal amplification
with the filter on. The recorded signal must be less than 4 mm.
2014.01.09
Heart Screen 60G-80G-L
P2
Examined lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
P2
Examined lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
Measured amplitude
Measured amplitude
36/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
Service manual