HeartScreen 60G Basic functions

1.
Connect the keyboard, the LCD and the generator.
2.
Before downloading the program connect a power supply set to +9V in place of the battery. Remove the resistor.
3.
Connect the other power supply – set to 3.6V – to the J17 battery drain point. Connect the negative points of the
two power supplies.
Test procedure
Keyboard test
Speaker test
RTC (clock) test
Contrast test
Serial data transfer test
Serial data transfer test with SIO
reducing cable
Low Batt. check
Checking power off
Concerning measurement procedures not controlled above, the IEC601/1 safety and IEC 601-2-25 ECG device related
standards are authoritative.
8.3

Checking technical requirements

8.3.1 Safety examinations

In case of repeated examinations the high-voltage sparking examination has to be performed with decreased voltage.
In this case the voltages are at least 80% of the given value.

8.3.1.1 Insulation checking

High-voltage insulation checking has to be performed on the device with the following voltage values:
1.5 kVeff between the power supply's primary and secondary side
4 kVeff between the device's secondary side and the patient connection
4 kVeff between the charging adapter connection and the PC -SIO connection
4 kVeff between the patient connection and the PC-SIO connection
2014.01.09
Heart Screen 60G-80G-L
Comment
Make sure the appropriate function is performed when the buttons are
pressed.
During ECG signal monitoring, if the speaker is turned on.
Set the exact date and time, which has to be the same after turning the card
off and on again.
The LCD contrast is acceptable, not foggy, the LCD is operational.
ECG signal monitoring with Innobase for Windows program and generator,
checking main functions.
ECG signal monitoring with Innobase for Windows program and generator.
Gradually decrease battery voltage and with it the ONPOWER voltage
level (the processor checks this every second), at 8.4 +/-0.1V a symbol
appears in the bottom right corner:
(ONPOWER J11.9)
Continue to decrease the battery voltage (the processor checks this every
second), the device turns off at 7.2 +/-0,1V (ONPOWER J11.9).