a/ In case of DC = 0
The signal's amplitude on the measured lead should be 16000RV +/-320RV (+/- 2 %).
P1
L
R
V1
V2
V3
V4
V5
V6
b/ In case of DC = -320 mV
The signal's amplitude on the measured lead should be 16000RV +/-320RV (+/- 2 %).
P1
L
R
V1
V2
V3
V4
V5
V6
c/ In case of DC = +320 mV
The signal's amplitude on the measured lead should be 16000RV +/-320RV (+/- 2 %).
P1
L
R
V1
V2
V3
V4
V5
V6
2014.01.09
Heart Screen 60G-80G-L
P2
Measured lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
P2
Measured lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
P2
Measured lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6