6.5.6 Checking time-constant

Apply (8000 +/-0.5% RVpp / 0.05 Hz sine) + DC signal on the amplifier input according to the tables from a ground-
independent generator. DC should be +320, -320 mV. Set the device to single sensitivity (x1) and 5 mm/sec speed.
Filter is off.
Connect the patient cable according to the following tables, start the individual channels with the channel selector and
measure the signal amplitude. The time-constant should be at least 3.2sec:
a/ In case of DC = -320 mV
P1
L
R
V1
V2
V3
V4
V5
V6
b/ In case of DC= +320 mV
P1
L
R
V1
V2
V3
V4
V5
V6
2014.01.09
Heart Screen 60G-80G-L
P2
Measured lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
P2
Measured lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
P2
Measured lead
all other
I
all other
II
all other
V1
all other
V2
all other
V3
all other
V4
all other
V5
all other
V6
27/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
Service manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heart screen 80g-l

Table of Contents