User Manual MDCEN-302-SMD Microprocessor Centrifuge PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.7.22...
Page 2
MDCEN-302-SMD Instruction Manual for Microprocessor Centrifuge (DC brushless motor) Emergency device Emergency bar Main power switch Warnings: * Do not open the lid before the centrifuge stops completely. * Do not remove the centrifuge during operation. * Do not put objects on the top of the lid.
Page 3
the button . The icon RCF will appear on the screen, and the RCF value is flash on the screen. Push the button to set RCF value. Push the button or any other function button to enter the RCF value. 6.
Page 4
auto brake function (which means when the machine stop rotating, there is no auto brake function, and the machine will stop rotating naturally.) Push the button continuously, and at the same time, push the button ,the BRK will disappear. Repeat the same step, the auto brake function can restore.
Page 5
1. Keep the centrifuge clean all the time. Cleaning can be done with a damp cloth. Avoid the use of solvents as they may damage the product housing, rotor. 2. Check if the rotor is in the center. In case, the rotor is slanted. Stop using machine and contact the distributor for repair.
Page 6
Service Manual Troubleshooting Guide Caution: Always disconnect the power cord before troubleshooting. Trouble / error message Cause Remedy Turn the main power switch off, and Motor is not working then turn it on again. 1. Rotor type setting is not correct. 1.
Page 7
4. PC board is defective. 4. Contract the distributor for repair. 1. Fuse is wasted 1. Replace a new fuse. Burnt fuse 2. Switching power is defective. 2. Contract the distributor for repair. Lid can’t be open 1. Lid lock is defective. 1.