Check the 4 screws that connect the interchangeable fitting to the
measuring block for a solid seating using an 8 mm Allen key.
Remove the two-piece hood from the electronic sensor assembly.
Remove the protection cap for transportation from the sensor element.
Turn the interchangeable fitting using only your hand into the position
CLOSE. This means that the interchangeable fitting seals off the interior
of the pipeline.
Insert the electronic sensor assembly without sensor protection cap into
the interchangeable fitting. Ensure the correct alignment. The electronic
Check tightness of
the 4 screws that
fasten the inter-
changeable fitting
Alignment pins for hood (2-
part)
Hood (2-part)
Protection cap for trans-
portation for sensor tip
(included in delivery)