5 Installation
5.5
Pulling direction for remov-
ing the hood parts
26
Installation of the electronic sensor
assembly in the interchangeable
fitting (version 0699 6447)
•
Check the 4 screws that connect the interchangeable fitting to the
measuring block for a solid seating using an 8 mm Allen key.
•
Remove the two-piece hood from the electronic sensor assembly.
•
Remove the protection cap for transportation from the sensor element.
•
Turn the interchangeable fitting using only your hand into the position
CLOSE. This means that the interchangeable fitting seals off the interior
of the pipeline.
•
Insert the electronic sensor assembly without sensor protection cap into
the interchangeable fitting. Ensure the correct alignment. The electronic
Check tightness of
the 4 screws that
fasten the inter-
changeable fitting
Alignment pins for hood (2-
part)
Hood (2-part)
Protection cap for trans-
portation for sensor tip
(included in delivery)