6.

Operating the CV5030 in connection with the TS5025 transfer station

6.1
Differences compared to CV 5030 stand-alone operation
6.2

Requests by the robotic color multistainer ST5020

30
If the TS5025 is connected, the message "TS 5025 PRESENT" appears
for 1 s on the display after activating the main power switch.
• Requests by ST5020 are accepted in READY mode (request approx.
every 5-10 seconds).
Prerequisite:
• Use the insert for Leica slide racks (30) in the bath.
• Keep the bath clear, i.e. do not deposit any slide racks.
• Refill consumables (coverslips, mountant) as soon as possible.
• Empty output racks as soon as possible.
• Empty slide racks must be removed from the output slide ramp
of the TS5025. The second deposited rack activates a sensor that
prevents the processing of additional racks.
• Requests are permitted if the CV 5030 is in READY mode (or a rack is
being processed) and no interaction is required of the user (e.g.
priming, emptying of output magazine):
• The slide rack is transferred by ST5020.
• After all slides have been coverslipped (for interruptions see sec-
tion 5.5), the empty slide rack is discarded via the ramp of the
TS5025 transfer station.
• If the output racks are full or the slide rack is already the third on
the ramp, the user is requested to remove it.
• Status changes to READY.
• ST5020 recognizes if the CV5030 is switched off or not switched to
standby mode (no feedback). If the CV5030 is not ready for operation,
the racks are placed in the exit station of the ST5020 and must be
transferred manually.
Instruction Manual V1.1 - 11/2002