36
2.11
Protective conductor test
SIE
Protective conductor test
Observe the information on the protective conductor resistance test contained in these
instructions.
•
The protective conductor test is to be carried out according to ARTD-002.731.17... with
the unit closed.
The protective conductor resistance must not exceed 0.2 Ohm.
•
The determined values are to be logged and evaluated in the report "Protective conduc-
tor resistance / report" by indicating the measuring points.
•
In addition, the measuring method and the measuring device used (designation and se-
rial number) are to be documented.
NOTE
NOTE
POLYMOBIL Plus
The report "Protective conductor resistance / report" is filed in the
"Certificates" register in the system folder.
If documentation of protective conductor resistance measure-
ments is missing, a new report must be created.
A new report "Protective conductor resistance / report" is included
in the SPR8-125.034.02.02 startup instructions. Detach this report
from these instructions, complete it and file it in the "Certificates"
register in the system folder. For a first determination, the deter-
mined values are to be logged as values measured for the first
time. Evaluation is not required in this case.
SPR8-125.831.01.02.02
05.05
Inspection and Maintenance
Page 36 of 40
CS PS 24
Siemens AG
Medical Solutions