Inspection and Maintenance
2.6
Single tank
Fig. 2:
SIM
Suspension
•
The single tank support
on the flange of the support arm. Check the torque of the 4 screws
•
T
NOM: 25 Nm, Tolerance: ± 10%.
SIM
Locking device
•
Check the locking handle
•
With the locking device released (locking handle
wise), it must be possible to move the single tank to all sides easily.
•
Checking the locking device: Tighten the locking handle
direction. It should no longer be possible to shift the single tank.
Readjusting the locking device:
- Tighten the locking handle
- Unscrew the screw
the screwdriver.
- Insert the locking handle in perpendicular position and tighten the screw
(4//Fig. 3 / p. 27)
- Recheck the locking mechanism of the single tank.
Siemens AG
Medical Solutions
(1//Fig. 2 / p. 27)
(3//Fig. 3 / p. 27)
(3//Fig. 3 / p. 27)
(4//Fig. 3 / p. 27)
securely.
SPR8-125.831.01.02.02
05.05
Fig. 3:
with the single tank must be seated securely
for firm seating and tighten, if necessary.
(3//Fig. 3 / p. 27)
(3//Fig. 3 / p. 27)
in clockwise direction up to the stop.
and push the locking handle off the shaft using
Page 27 of 40
CS PS 24
27
(2//Fig. 2 / p.
27);
set counterclock-
in clockwise
POLYMOBIL Plus