Double-Slot Diaphragm - Siemens POLYMOBIL Plus Maintenance Instruction

Hide thumbs Also See for POLYMOBIL Plus:
Table of Contents

Advertisement

28
2.7

Double-slot diaphragm

Fig. 4:
Connect the line voltage cable and turn the Polymobil Plus ON .
PMF
Light localizer lamp
Check the function of the light localizer lamp and replace, if necessary. Refer to the Ser-
vice Instructions "Replacing the light localizer lamp".
SIM
Rotating
Check whether the double-slot diaphragm can easily be rotated ± 45
PMF
Format setting
Check the format setting (pairs of collimator leaves) for ease of movement and unifor-
mity (by sliding the diaphragm adjusting levers
Switch the Polymobil Plus OFF and disconnect the line voltage plug.
PMF
0° degree position
Move the support arm into its uppermost position.
Place a water level against the double-slot collimator and check the 0° setting.
If necessary, readjust the scale on both sides.
POLYMOBIL Plus
SPR8-125.831.01.02.02
05.05
Inspection and Maintenance
(1//Fig. 4 / p.
Page 28 of 40
CS PS 24
(Fig. 4 / p.
28).
28).
Siemens AG
Medical Solutions