Specifications; Certification - UEi C163KIT Instruction Manual

Residential / commercial combustion analyzer with heat exchanger test
Table of Contents

Advertisement

Parameter
Temperature Measurement
Flue Temperature
Inlet Temperature (Internal Sensor)
Inlet Temperature (External Sensor)
Flue Gas Measurement
Carbon Monoxide
Calculations
Oxygen
CO/CO
Ratio
2
Efficiency (Net or Gross)
Efficiency High (C)
Excess Air
Pressure (Differential)
Pre-programmed Fuels
Battery Life

Certification

Operating Conditions
Temperatures
Humidity
Power Supply
Physical Characteristics
Weight
Dimensions
These analyzers are in conformity with the relevant Union harmonization legislation listed below:
Directive
201430EU
201165EU
The C163 are TUV-tested and certified to EN 50379, Parts 1, 2 & 3 in accordance to 1st German Federal Emission Control Ordinance
(BlmSchV)
EMC
EN507270:2015
Safety
EN61010-1:2010
RoHS
IEC62321-2:2013, IEC62321-1:2013; IEC62321-3-1;2013, IEC63321-5:2013, IEC623321-4:2013, IEC62321-7-2:2017, IEC62321-7-1:2015,
IEC62321-6:2015
AHRI
1260
18
Range
32˚ to 1112˚F (0˚ to 600˚)
32˚ to 122˚F (0° to 50°C)
32˚ to 1112˚F (0˚ to 600˚)
0 - 2000ppm
0 - 21%
0 - 0.9999
0 - 99.9%
0 - 119.9%
0 - 119.9%
±80mbar
Nat Gas, H Oil, Pelllets, Oil, LPG, Butane, Propane
>8 hours (continuous with pump on)
The C162/C163/C164 are TUV-tested and certified to EN 50379, Parts 1-3 in accordance to 1st
German Federal Emission Control Ordinance (Bim5chV); Meets requirements for AHRI 1260
32˚ to 113˚F (0˚ to 45˚C)
15 to 90% RH, (non-condensing)
Rechargeable batteries, USB Charging
1.5 lb. (680g)
H: 8.54"x W: 4.18"x D: 1.86" (H: 217 mm x W:106 mm x D: 47mm)
Electromagnetic Compatibility (EMC)
Restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS)

SPECIFICATIONS

Resolution
0.1˚F (0.1˚C)
0.2˚F (0.1˚C)
0.2˚ (0.1˚C)
1ppm
0.1%
0.0001
0.1%
0.1%
0.1%
0.1mbar
Title
CERTIFICATION
Accuracy
±(0 .3% rdg +3 .6˚F (2˚C))
±2˚F (1˚C)
±2˚F (1˚C)
±3ppm or ±5% rdg
(whichever is greater)
±0.3% volume
±5% rdg
±1% rdg
±1% rdg
±0.2% rdg
±0.5% FSD

Advertisement

Table of Contents
loading

This manual is also suitable for:

C163

Table of Contents