6. Image quality performance 2
・Negative film reproducible density range
 Under conditions of CMS off, γ=1, and 16 bits, output images should be scanned within the range from
the min. value is (0) to max. value (65535) of RGB.
・Gradation sequence
The film with up to 4.2 density should be scanned accurately within ±0.1 of the difference between the
output density and ideal density.
Indistinctive tone jump
 
・Density expressible range:
 4.2 or more
・Prescan reproducibility
 The color variation of 24-patch Lab should be ΔE4 or less. (Reference value: average of 5-times
scanning)
・Drift
Just after power is applied and after operation lasts 30 minutes, the color difference ΔE of the scanned
image of the same chart should be 5 or less on average and 10 or less at a maximum. The shading
standard should be met.
・Streaking
 No streaks in both main- and sub-scanning directions.
 Not as distinctive as LS-8000
・Grain roughness
 Indistinctive. No significant difference from the traditional models.
・Irregular color
 Color difference (ΔE) should be 5 or less.
・Influence of outside light
 No influence
7. Operating sounds
 No significant abnormal sounds with 55-dB or less at a 1-m distance away.
8. Status indication
0.635Hz ± 1% of blinking frequency
 BUSY:
5Hz ± 1% of blinking frequency
 ERROR:
- R4 ・
-
LS-9000
VRA51701-R.3607.A