4.3 Upgrade of TS2000 Software

4.4 Additional Software Installation

SPB7-520.840.01.01.02
On a time-to-time basis, TransScan will supply a new version of
software for the
TS2000 System. The upgrade will be supplied on a CD-ROM, and
will be performed by the service engineer.
To perform the upgrade:
Install the CD with the new software version into the CD-
1.
ROM Drive (the CD has an auto-run program).
Wait for approximately 15 seconds after the program
2.
begins and click "Next."
After successfully loading the software, click on
3.
"Finish".
If AutoRun does not start, proceed as follows:
From the "START" Menu select "RUN", then
1.
"BROWSE".
From the CD Menu select "SETUP"/ "OPEN".
2.
Click "OK"; the program will then install the new
3.
version on your system.
Click "FINISH", remove the CD and restart the system.
4.
To install the software used in this system (which should be in the
"windows" screen), follow the instructions below:

4.4.1 WIN NT4 Installation (Including Service Pack 4)

Access the computer's BIOS setup. (Press "Del" when
1.
the computer boots-up.)
From "Boot Order," select CD-ROM as the first boot
2.
drive.
Insert the WIN NT4 CD into the CD-ROM drive.
3.
Re-boot the system. The WIN NT4 installation begins
4.
automatically.
Select "Setup Windows NT Now," and press "Enter"
5.
twice.
Press "Page Down" and then press F8.
6.
RD7409290-0 Rev. F
4-3