Thank you for purchasing "Cold Trap, CA Series"
of Yamato Scientific Co., Ltd.
To use this unit properly, read this "Instruction
Manual" thoroughly before using this unit.
Keep this instruction manual around this unit for
referring at anytime.
WARNING!:
Carefully read and thoroughly understand the
important warning items described in this manual
before using this unit.
Yamato Scientific Co. LTD.

Advertisement

Table of Contents
loading

Summary of Contents for Yamato CA300

  • Page 1 Instruction Manual - First Edition - Thank you for purchasing "Cold Trap, CA Series" of Yamato Scientific Co., Ltd. To use this unit properly, read this "Instruction Manual" thoroughly before using this unit. Keep this instruction manual around this unit for referring at anytime.
  • Page 3: Table Of Contents

    Contents Cautions in Using with Safety..............1 • Explanation........................1 • Table of Illustrated Symbols ..................2 • Fundamental Matters of “WARNING!” and “CAUTION!” ..........3 Before Using this unit ................5 • Requirements for Installation..................5 Description and Function of Each Part ..........8 •...
  • Page 5: Cautions In Using With Safety

    Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLS Illustrated Symbols Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below.Be sure that you understand the warnings and cautions in this manual before operating the unit.
  • Page 6: Table Of Illustrated Symbols

    Cautions in Using with Safety Table of Illustrated Symbols Warning Warning, Warning, Warning, Warning, Warning, generally high voltage high temperature drive train explosive Caution Caution, Caution, Caution, Caution, Caution, generally electrical shock scald no road heating not to drench Caution, Caution, water only deadly poison...
  • Page 7: Fundamental Matters Of "Warning!" And "Caution

    Cautions in Using with Safety Fundamental Matters of “WARNING!” and “CAUTION!” WARNING! Do not use this unit in an area where there is flammable or explosive gas Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof.
  • Page 8 Cautions in Using with Safety Fundamental Matters of “WARNING!” and “CAUTION!” CAUTION! During a thunder storm During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and the main power. If this procedure is not followed, fire or electrical shock may be caused. Do not touch the condensed liquid in the cooling coil and trap bath Since the condensed liquid in the cooling coil and trap bath is stayed with low temperature, never touch it so as to preventing from getting frostbite on your hands.
  • Page 9: Before Using This Unit

    • Do not install this unit in a place where: ♦ Rough or dirty surface. ♦ Flammable gas or corrosive gas is generated. ♦ Ambient temperature bellow 5℃ or above 35°C. (CA300) ♦ Ambient temperature bellow 5℃ or above 30°C. (CA800) ♦ Ambient temperature fluctuates violently.
  • Page 10 Before Using this unit Requirements for Installation 4. Do not modify • Modification of this unit is strictly prohibited. This could cause a failure. 5. Do not topple or tilt this unit • Set this unit to the flattest place. This unit incorporates the refrigerator.
  • Page 11 • Choose a correct power distribution board or receptacle that meets the unit’s rated electric capacity. Electric capacity: CA300: 100V AC, 50/60Hz, 6A CA800: 100V AC, 50/60Hz, 10A NOTE) Starburst connection with a branching receptacle or extended wiring with a cord reel lowers electrical power voltage, which may cause the degradation of refrigeration capability.
  • Page 12: Description And Function Of Each Part

    Description and Function of Each Part Main Unit and Control panel Main unit Cover with nozzle Production plate Control panel Specification plate Power switch (Earth leakage breaker) Ventilation port Drain hose (The drain hose of CA800 is installed on the right side of Caster the body.) Power code...
  • Page 13: Installation Method

    Installation Method 1. Unlock the stopper of the caster. Pulling up the lever of the stopper for caster releases the lock. (Only the two casters in front of the unit are attached the stopped.) 2. Move the device to the place to be Release lock installed.
  • Page 14: Operation Method

    Operation Method Procedure of Operation Read through the following procedure if the liquid to be trapped is water type. Use the glass condenser set (optional accessory) for the acid or organic solvents. Refer to the operation manual attached to the glass condenser for using this glass condenser. Supply the power 1.
  • Page 15: Cooling Curve (Reference Data)

    Operation Method Cooling Curve (Reference Data) CA300 ■ Ethanol 3liters Under: 100V, 50Hz Room temp: 20℃ Liquid temperature with three liters of ethanol in trap 0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 Time(min.)...
  • Page 16 Operation Method Cooling Curve (Reference Data) CA800 ■ Ethanol 3liters Under: 100V, 50Hz Room temp: 20℃ Liquid temperature with three liters of ethanol in trap 0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 Time (min.) ■...
  • Page 17: Handling Precautions

    CAUTION! Maximum trap capacity The max. trap capacity for the water type liquid is around 0.9kg (CA300) or 0.8kg (CA800). Inserting too much trap to this unit could cause the clogging of "IN", "OUT" nozzle. Do not step on this unit Do not step on this unit.
  • Page 18: Maintenance Method

    Maintenance Method Daily Inspection and Maintenance WARNING! • Disconnect the power cable from the power source when doing an inspection or maintenance unless needed. • Perform the daily inspection and maintenance after returning the temperature of this unit to the normal one.
  • Page 19: Long Storage And Disposal

    Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… • Turn off the power and disconnect the power cord. WARNING! When disposing… • Keep out of reach of children. •...
  • Page 20: In The Event Of Failure

    In case of lighting the alarm lamp again, make a call to the service office. In the case if the error other than listed above occurred, turn off the power switch and primary power source immediately. Contact the shop of your purchase or nearest Yamato Scientific Service Office.
  • Page 21: After Service And Warranty

    In Case of Request for Repair If the failure occurs, stop the operation, turn OFF the power switch, and unplug the power plug. Please contact the sales agency that this unit was purchased, or the Yamato Scientific's sales office. < Check following items before contact >...
  • Page 22: Specification

    W284×D404×H700 mm W405×D455×H850 mm Weight Approx. 32Kg Approx. 54Kg Power supply 100V AC, 50/60Hz, 6A 100V AC, 50/60Hz, 10A Accessory Instruction manual Optional Accessories (common for CA300/800) Name Product code Notes Collection acid organic Glass condenser set OCA10 2221487 solvents Brass φ30×φ18...
  • Page 23: Wiring Diagram

    Wiring Diagram CA300 青 赤 AC100V ③ ④ 茶 ⑩ ① 1 ⑤ ⑪ ② 2 X1-b 3 ⑧ 4 ① ⑥ ⑦ Symbol Part name Symbol Part name Earth leakage breaker Pressure switch Refrigerator switch Operation condenser (FM) Terminal block...
  • Page 24 Wiring Diagram CA800 AC100V ⑥ ① 1 ② 2 ③ ⑩ ⑪ ⑨ 3 ⑯ X4-2 OVR2 ⑬ ⑫ 4 ⑮ ③ ⑤ 5 ⑭ ⑦ ⑧ X4-1 OVR1 Symbol Part name Symbol Part name Earth leakage breaker OVR1, OVR2 Overload relay Refrigerator switch C1, C3...
  • Page 25: Replacement Parts Table

    Replacement Parts Table CA300 Part Name Code No. Specification Manufacturer Packing CA300-30150 Silicon Yamato Scientific Refrigerator unit 3-01-003-6005 100V/300W UF-NS300L-YL Yamato Scientific Inner dia. φ 0.8 × outer dia. φ 1.8 Yamato Scientific Capillary 3-07-002-0004 Refrigerator switch 2-01-027-0005 HLN215A Fujisoku...
  • Page 26: Reference

    Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable ingredients in this unit. EXPLOSIVE Ethylene glycol dinitrate (nitro glycol), Glycerin trinitrate (nitroglycerine), Cellulose nitrate (nitrocellulose), and other explosive nitrate esters Trinitrobenzene, Trinitrotoluene, Trinitrophenol (picric acid), and other explosive EXPLOSIVE: nitro compounds Acetyl hidroperoxide (peracetic acid), Methyl ethyl ketone peroxide, Benzyl...
  • Page 27 Responsibility Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents or breakdown of device if it is used with a failure to comply. Never conduct what this document forbids. Unexpected accidents or breakdown may result in.