Ultrasonic Pipette Washer Model AW31 Instruction Manual First Edition ● Thank you for choosing the AW31 Ultrasonic Pipette Washer by Yamato Scientific Co., Ltd. ● For proper equipment operation, please read this instruction manual thoroughly before use. Always keep equipment documentation safe and close at hand for convenient future reference.
TABLE OF CONTENTS 1. SAFETY PRECAUTIONS ......................... 1 Explanation of Symbols........................1 Warnings and Cautions ........................2 2. COMPONENT NAMES AND LOCATIONS..................5 Main Unit ............................5 3. CLEANSING PROCESS........................6 Process Summary..........................6 4. OPERATION PROCEDURE......................7 Preparation............................7 Operation Steps..........................
1. SAFETY PRECAUTIONS Explanation of Symbols A Word Regarding Symbols Various symbols are provided throug hout this t ext and on equ ipment t o ensure safe o peration. F ailure to comprehend t he operational hazards and risk s associated with these symbols ma y lead to adve rs e results a s explained below.
Place in a location with adequate drainage Do not use AW31 where poor drainage may cause drain water to backup into and flood unit, possibly resulting fire or electric shock hazard. DO NOT allow equipment to become wet ・ ...
Page 6
A direct lightning strike may cause damage to equipment, or result in fire or electric shock. Equipment damage/breakage If AW31 unit tips over or falls and is damaged, or if wash container breaks, turn off main switch and disconnect power cord immediately. Contact a local dealer or Yamato sales office for assistance.
Page 7
Do not plug or unplug the power cable with wet hands. Doing so may result in electric shock. Humidity and dust Do not use AW31 unit in overly humid or dusty locations. Doing so my result in fire or electric shock. Extended out-of-service periods If unit is to be stored or put out of service for an extended period, be sure to turn off the power switch and disconnect the power cord.
2. COMPONENT NAMES AND LOCATIONS Main Unit AW31 Unit Overview Container Lid Cage Alignment Pin Wash Containter Pipette Cage Drain Pipe Power Cable Unit Base Power Indicator Lamp Rotary Shaft Power Switch Drain Outlet Water Supply Inlet 5 ...
The Wash Cycle The AW31 utilizes ultrasonic waves, emitted from the bottom surface of unit, to break up and free contaminants from glass pipettes as they soak in a detergent bath. A fluid jet stream is also applied to assist in freeing contaminants left on inner and outer pipette surfaces. To ensure an even and uniform wash, pipettes are rotated in a cage while the ultrasonic waves and fluid jet are applied.
1. Setup Location Warning Set up AW31 unit in a location with good drainage. Do not place unit where fluids can splash onto outside of unit or where drainage is insufficient, causing fluid to back up and overflow onto base unit.
4. OPERATION PROCEDURE Operation Steps Washing 1. Fill wash container with water to approximately 50cm deep (about 10ℓ) and add the specified amount of detergent. 2. Insert pipettes and place lid on wash container. Be sure pipette cage is properly aligned and engaged on rotation shaft and that cage alignment pin on lid is properly centered in pipette cage when covering wash container.
Proper Pipette Handling & Cleansing To get the most out of the AW31 unit’s cleansing potential, be sure to follow the steps below: 1. Put pipettes into water immediately after using and do not allow them to dry out before washing. If pipettes are allowed to dry with contaminants or chemicals on them, they become more difficult to clean.
1. Do not allow air to collect in the drain hose. Fluid will not siphon off and cycle through properly if air becomes trapped in drain lines. 2. The AW31 unit is rated for use at 100V AC. In the event that voltage drops below 90V, machine performance may lag. Conversely, if voltage exceeds 110V, equipment damage may result. Confirm voltage rating before use and be sure that power supply is not susceptible to extreme fluctuations in voltage.
7. TROUBLESHOOTING Troubleshooting Guide Symptom Possible Causes Possible Solutions a) Water level in wash container Fill wash container to over 1. Power indicator lamp comes is less than 10cm deep. 10cm deep. on, but pump and oscillator do b) Purified or distilled water is Use tap water.
10. REPLACEMENT PARTS TABLE Below is a list of replacement parts which can be provided upon request. Description Part Number Specification Number Wash Container Lid W000000118 GFTA0024UCZZ Cage Alignment Pin W000000127 LPIN-1003UCZZ Wash Container Outer Tube W000000130 LSOU-0023UCZZ Cage Unit W000000133 NKAG-0002UCZZ Cage Base...
Page 18
Notice ●Instruction manual descriptions and specifications are subject to change without notice. ● Yamato Scientific Co., Ltd. will replace flawed instruction manuals (pages missing, pages out of order, etc.) upon request. Instruction Manual...
Need help?
Do you have a question about the AW31 and is the answer not in the manual?