9 - Click Mode Operating Instructions
(Option)
9.1 Introduction
The CLICK mode allows the User to make in a fast, easy and fully
automatic way a difficult test like the Discontinuous Disturbances
measurement, as defined by current CISPR standards.
To enter in Click Mode press the key in the main menu, and immediately
the first window opens as follow:
Fig. 9-1 Click
The Click mode function has four commands:
• Start
• Clear: cancel and reset all data to start an idle test;
• Report
• SETUP
Always use Esc button to return to the previous view/condition.
To increase test productivity the PMM 9010 has some very unique features:
it allows skipping as many as 2 steps, as it may predict how many clicks
would overcome the Lq limit and may also advice the User if the next step
would fail.
Moreover, it makes a Qpeak measure and not a mere threshold
comparison, as only a fully digital receiver can detect whether a Qpeak
level came from an unsaturated signal or not: an analogue analyzer only
have the final weighted Qpeak.
Document 9010EN-70306-1.59 - © NARDA 2007
Click Mode Operating Instructions
9-1