PART FIVE
5. CIS SENSORS AND DETECTION LEVEL ADJUSTMENT
1)

CIS sensor Calibration

Contact AMROTEC to purchase CIS Calibration sheet.
• How to insert CIS calibration sheet: (See photo illustration below)
a) Open the rear backdoor
b) Open the bill path (press Tab towards below) and clear CIS sensor with soft
cloth.
c) Insert calibration sheet all the way in though the opening of the Bill path
area until it fully blocks CIS sensor.
d) Close the bill path and Rear door
• Prepare to Calibration process
a) Press MENU key on the control panel to enter into MENU screen>Press START
Key or #1 key on the control panel to Enter into (1) Service Menu
b) Input password: 2552 using numeric keys on the control panel
c) On SERVICE MENU, scroll to (2) CIS calibration by using down arrow  or just
press [2] key on the control panel and then Press [START/STOP] Key.
d) By pressing the [START/STOP] key will start the CIS calibration process. The
display screen will show as below in sequence:
CIS Calibration sheet
11