Manual N 2 Gas (O 2 Gas) Supply Line A/B Changeover - Panasonic MCO-170M Series Operating Instructions Manual

O2/co2 incubator
Table of Contents

Advertisement

Manual N
gas (O
2
It is possible to change N
Example) Change N
/O
2
1. Press N
/O
gas supply line select key for a few
2
2
seconds.
2. N
/O
gas supply line A is changed to B.
2
2
Note: The behavior for the following case is shown in Table 3.
After the N
/O
gas supply line A is changed to B by N
2
2
gas supply line B is changed to A manually without pressing the Buzzer key.
Table 3
Situation
Supply line Cylinder A Cylinder B
N
/O
gas supply line
2
2
A is changed to B
1
automatically.
Without pressing the
Buzzer key,
long-pressed N
/O
2
2
2
gas supply line select
key.
gas) supply line A/B changeover
2
/O
gas supply line A and B manually anytime.
2
2
gas supply line A to B.
2
N
gas
2
B
Empty
Remaining
A
Empty
Remaining
gas (O
gas) automatic changer function, the N
2
2
N
/O
gas supply line indicator
2
2
A
Reverse
video
and blink
Blink
/O
2
2
Message
display field
B
Err02:
Light on
N2/O2 Gas Empty
(and buzzer)
Err02:
Light off
N2/O2 Gas Empty
(and buzzer)