OPERATION/ALARM LOG
4. On the Setting screen, input Log Interval. Press
Apply key to save the input value. The display
returns to the Log screen.
Settable range: 2 minutes~30 minutes.
Factory setting: 6 minutes.
Note: It is possible to register 8-digit alphanumeric
characters as the Unique ID. Refer to page 56.
Note: Relation between log interval and the estimated amount of data that can be saved
Log interval=2 minutes: Approx. 46 days
Log interval=6 minutes: Approx. 135 days
Log interval=30 minutes: Approx. 664 days
When saving data more than the above, and the data is overwritten and the old data is delated.
5. Press Top key to return to the Top screen.

Displaying operation log

Operation log saved in the incubator can be displayed graphically on the LCD touch panel.
1. Press Menu key to lead the Menu screen.
2. Press Log key to lead the Log screen.