Panasonic MCO-170M Series Operating Instructions Manual page 75

O2/co2 incubator
Table of Contents

Advertisement

H
O
DECONTAMINATION
2
2
●Step4 is the process to decontaminate in the
chamber by generating H
generator.
Note:
By
pressing
decontamination is stopped in the middle of
decontamination and go to Step7 (The process to
resolve H
O
vapor by UV lamp). It is not possible
2
2
to skip Step7.
●Step7 is the process to resolve H
lamp.
Note: Step5 and Step6 do not exist.
15. When the H
O
2
2
the
H
O
Decontamination
2
2
displayed.
Open the outer and inner doors. Disconnect
connector of the chamber, cable and H
generator, and remove H
from the chamber.
Note: When doing the above work, put on
protection glasses and rubber gloves.
16. On the H
O
Decontamination Step8 screen,
2
2
press OK key to return to the Top screen.
Note: When H
O
2
2
connector of the chamber by cable, OK key is not
workable.
17. Dilute the remaining H
Rinse and wash the H
environment outside of the chamber.
Note: Do not wash either the inside or outside of the H
74
O
vapor from H
2
2
Abort
key,
O
vapor by UV
2
2
decontamination is completed,
Step8 screen
O
generator and cable
2
2
generator is connected with
O
reagent in the H
2
2
O
generator with distilled water. Then keep the H
2
2
O
2
2
H
O
2
2
is
O
2
2
O
generator with a large volume of water and dispose of it.
2
2
O
generator with alcohol.
2
2
O
generator in a clean
2
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-170muv seriesMco-170muvh series

Table of Contents