Filling humidifying pan

Use the following procedure to fill the humidifying pan with water or to replace water in the humidifying
pan.
1. Pull out the humidifying pan toward you. (Fig. 1)
2. Dispose of the remaining water in the humidifying
pan and clean the humidifying pan with a diluted
detergent. Then rinse it thoroughly with distilled
water and wipe it with alcohol for disinfection.
3. Wipe all moisture from the bottom of the chamber.
4. Return the humidifying pan to the chamber and
pour sterile distilled water (approx. 1.5 L, preheated
to 37 C). (Fig. 2)
Note:
・Operation with no humidifying water may increase the chamber temperature than the set temperature
temporarily.
・Preheat the water to 37 C to be poured into the humidifying pan. Adding low-temperature water will lower
the temperature and humidity in the chamber.
・Install the humidifying pan in a longitudinal direction as its shorter side is placed in the back.
・Refill the humidifying pan with water early when the volume of water is decreased.
・Mixing any reagent in the humidifying water may have a bad influence on the cultivation. Especially when
using the UV lamp, do not use any reagent. Because the UV light may deteriorate the reagent mixed with
the humidifying water.
5. Set the humidifying pan with the inner side flush against the back, and close the inner door and the outer
door.
Note: Set the humidifying pan with the inner side flush against the back. The humidity control bar in the
duct keeps at low temperatures and inner moisture is recondensed. Slide the humidifying pan down right
under the humidity control bar, otherwise the recondensed water drops will directly fall to the chamber
bottom and will pool in the chamber bottom.
When the pooled water evaporates, it may leave a white mark on the chamber bottom. This is not a
malfunction. Wipe it off with alcohol-soaked gauze or unwoven cloth. When the mark cannot be removed,
scrub the mark off with using a cream cleanser.