18. After H
O
decontamination, surplus H
2
2
bottom of the chamber and in the bottom part of the H
duct. This solution contains H
glasses and rubber gloves, and wipe it up with a non-woven cloth.
19. Ventilate the chamber sufficiently and place all the attachments
back into the chamber.
Note: After H
O
decontamination, cover the connector on the
2
2
chamber side with the connector cap deeply. (Fig. 4)

Precautions when handling H

Observe the following precautions when handling optional H
 Handling precautions
・Wear protective equipment, such as protective glasses and rubber gloves.
・Do not use fire in the area where the reagent is being handled.
・Do not leave any reagent in the container after it has been used or while it is being used.
・Do not place inflammable or combustible materials near the area where the reagent is handled.
 Precautions for storage
・Store in a cool, dark place.
・Always close the container cover securely to prevent impurities from becoming mixed in the reagent.
・Check the container to make sure that there is no damage, corrosion, or cracking.
・Store the container with the inlet facing upwards, and make sure that the container will not tip or be
knocked over.
 Precautions for disposal
・Dispose according to the rule in your country.

Using unlock key

 Unlocking when power is interrupted
MCO-170MUVH or when an optional electric lock MCO-170EL is installed to the MCO-170MUV/170M,
outer door is locked with electric lock under a power outage. To unlock the outer door while the power is
interrupted, use the unlock key that is provided. To re-lock the outer door, turn the unlock key to the lock
direction while the outer door is open. After the outer door has been locked condition manually, then close
the outer door.
Note: The outer door cannot be locked by using the unlock key while the outer door is closed. Lock the
outer door while it is open. Attempting to turn the unlock key while the outer door is closed may damage
the electric lock system.
WARNING
Do not use the unlock key to unlock the outer door during H
gas resolve by UV. Doing so may cause harm to health from H
O
liquid will remain at the
2
2
O
at a low density. So put on protective
2
2
O
generator
2
2
O

reagent

2
2
O
reagent MCO-H2O2.
2
2
O
decontamination or during H
2
2
O
gas leakage.
2
2
Connector cap