Manual Co 2 Gas Supply Line Changeover - Panasonic MCO-170AICUVH Series Operating Instructions Manual

Co2

Advertisement

Manual CO
It is possible to change CO
Example) Change CO
1. Press CO
gas supply line select key for a few
2
seconds.
2. CO
gas supply line A is changed to B.
2
Note: The behavoir for the following case is shown in Table 3.
After the CO
gas supply line is changed over by CO
2
is manually retuned to the CO
Table 3
After changed into CO
changed into A manually.
Situation
Supply line Cylinder A Cylinder B
Changed into CO
2
gas supply line B
1
automatically
Not press Buzzer key,
long-pressed CO
2
2
gas supply line select
key
gas supply line changeover
2
gas supply line manually anytime.
2
gas supply line A to B.
2
gas supply line before the Buzzer key is pressed.
2
gas cylinder B because CO
2
CO
gas
2
B
Empty
Remaining
A
Empty
Remaining
gas automatic changer function, the empty CO
2
gas cylinder A is empty, if you
2
CO
gas supply line indicator
2
Reverse
video
and blink
Blink
gas
2
Message
A
B
display field
Err01:
Light on
CO
Gas Empty
2
(and buzzer)
Err01:
Light off
CO
Gas Empty
2
(and buzzer)
73

Advertisement

loading