Figure 12
Typical Partial Trace Confirming Trigger In Circuit Operation
6. Turn off the 8510.
7. On the 85102:
a. Remove the SMB-to-BNC cable.
b. For an 8360 series source, reconnect the Trigger In cable.
c. Reconnect W28 to A20J1.
d. Remove the SMB tee and reconnect W23 to A10/A12J5.
e. Replace the top cover.
f. For use with 8360 sources, place the BNC short over the SWEEP IN 0-10V BNC.
8. Turn on the 8510.
9. Press
INSTRUMENT STATE RECALL
10. Continue with
"Time Domain Option Functional Check"
domain option firmware (Option 010). If you did not install the time domain option
firmware, proceed to
85103E Performance Upgrade Installation Manual
,
[MORE], [FACTORY PRESET]
"Checking Calibration Kit Definitions" on page
85103E Upgrade Package (8510A to 8510C)
Checking the 8510C
.
if you have installed the time
31.
29