Loading the 8510C Operating System

Loading the 8510C Operating System
After you load the operating system, you will have an 8510C network analyzer, and it will
be referenced as such. Often the letter designation "C" at the end of the 8510C model
number is omitted.
The procedure for loading the operating system differs from the usual disk loading
procedure. This method allows you to load a disk with or without an operating system
previously loaded in the analyzer, and requires fewer keystrokes.
Selecting the appropriate 8510C operating system (CRT or LCD version) supplied in the
upgrade package will ensure that the proper operating system is loaded. The time domain
option firmware is included in both 8510C operating system versions and is enabled by the
time domain option security key IC.

Procedure

1. While holding down the
order:
a. Source (turn the switch to power on, not standby)
b. Test set
c. Analyzer
This causes the instrument to detect an apparent keyboard failure and display
self-test error 14, subtest 2. Ignore this error.
2. Press
again.
=MARKER
3. Insert the operating system disk into the analyzer disk drive, and press
4. Select the file named PG_8510C, then select
The disk loads in three to four minutes, and the system begins normal operation. The
screen displays: S11 Log Mag.
5. Press
INSTRUMENT STATE RECALL
6. Continue with
"Checking the 8510C."