Cleaning

1. Use "non-fluffing", non-abrasive, lint-free tissue or absorbent cloth. Hold the Piston
2. To remove all traces of contamination, the piston can be cleaned in a suitable solvent.
3. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure
4. Place piston carefully on a NEW tissue where it will not be damaged while the
4-2
Find Quality Products Online at:
O RING
Figure 4-1. Piston/Cylinder Removal
by the larger "head" end, and rub the tissue back and forth along its length.
O-ring seals should not be immersed in solvents, as they will
become damaged. They should be wiped carefully with a new
tissue.
above.
cylinder is cleaned.