Cleaning; Piston/Cylinder Removal - Fluke P3860-MPA User Manual

High pressure hydraulic deadweight tester
Hide thumbs Also See for P3860-MPA:
Table of Contents

Advertisement

P3800 Series
Users Manual

Cleaning

1. Use "non-fluffing", non-abrasive, lint-free tissue or absorbent cloth. Hold the Piston
by the larger "head" end, and rub the tissue back and forth along its length.
2. To remove all traces of contamination, the piston can be cleaned in a suitable solvent.
O-ring seals should not be immersed in solvents, as they will
become damaged. They should be wiped carefully with a new
tissue.
3. After removal from the solvent, using a NEW tissue, repeat the cleaning procedure
above.
4. Place piston carefully on a NEW tissue where it will not be damaged while the
cylinder is cleaned.
4-2
Shop for Fluke products online at:
O RING
Figure 4-1. Piston/Cylinder Removal
W Caution
MyFlukeStore
www.
PISTON NUT B
PISTON
PISTON NUT A
CYLINDER
PISTON COLUMN
1.800.561.8187
.ca
glf10.eps