D Set the prescale value.
.
ax" to "0
1. Set the prescale AX "ps
.
ay" to ". 10 00" and press the M [MODE] Key.
2. Set the prescale AY "ps
E Set the decimal point position.
1. Set the decimal point position "dp" to ",,,,," (default value) and press the
M [MODE] Key.
F Set comparative set value H to 700 and set comparative set value
L to 500.
1. Return to the RUN level by pressing the L [LEVEL] Key for at least 1 s.
(Start operation.)
2. Press the M [MODE] Key several times to change the SV display status to
"H" and set the value to "00700".
3. Press the M [MODE] Key several times to change the SV display status to
"L" and set the value to "00500".
G Start actual operation.
1. Press the M [MODE] Key several times to display the measurement values
and start actual operation.
Clearing Settings
If you become confused while setting the parameters and cannot
continue, all settings can be cleared so that you can start over.
Refer to "5.33 Initializing All Settings" (P.5-83) for information on
clearing all settings.
* Refer to "Section 5 Functions and Operations" for details on setting
parameters.
4.1 Initial Setup Example for the K3HB-R
.
1250" and press the M [MODE] Key.