Chapter 1, Calibrate the Test Set
Correcting for RF Path Loss
Procedure Overview
NOTE:
32
Using the RF Level Offset feature of the Test Set corrects for RF path power
losses that occur in cabling and connectors between the Test Set and a mobile
station.
It is highly recommended that RF path loss is corrected for. Besides returning
power measurements that do not have to be adjusted to account for RF path loss,
the Test Set uses the values you enter in this procedure to adjust the analyzer path
attenuation to an optimum value for demodulation and measurement functions.
The settings you made in this procedure must be re-entered after a power-cycle or
instrument preset.