4. Calibrate the Channel Power measurement.
1.Select Calibrate to start the calibration.
Calibration may take a minute or longer, depending on the RF Channel Std and Alt Chn Std field settings.
Average Power measurements are zeroed as part of the Channel Power calibration process.
Procedure Overview
Calibrating Channel Power measurements ensures that the Test Set achieves
accurate low-level CDMA power measurements.
Channel Power measurements:
Average Power measurements are zeroed as part of the Channel Power calibration
process.
indicate power in a 1.23 MHz bandwidth
are performed at one of the Test Set's IF frequencies by a DSP (Digital Signal
Processor)
use the Test Set's average power detector to calibrate the relative levels measured at IF.
Chapter 1, Calibrate the Test Set
Channel Power Measurement Calibration