Accuracy Specifications; Functional Specifications - Hioki PD3259-90 Instruction Manual

Digital phase detector
Table of Contents

Advertisement

Accuracy specifications
Conditions of
guaranteed accuracy
Line voltage
measurement accuracy
Frequency
measurement accuracy
Effects of external
magnetic fields
Effects of nearby wires
Temperature coefficient ±0.4 V/°C (at other than 23°C ±5°C)
Effects of humidity
Line voltage phase
difference shift
5.3 Functional Specifications
Phase detection
function
Guaranteed accuracy period: 1 year
Guaranteed accuracy period from adjustment
made by Hioki: 1 year
Temperature and humidity for guaranteed
accuracy: 23°C ±5°C (73°F ±9°F), 80% RH or
less
Warm-up time: Max. 10 sec.
Number of voltage sensor clip open/close
cycles: 8,000 or less
±2.0% rdg. ±8 dgt. (1-year accuracy)
±3.0% rdg. ±8 dgt. (3-year accuracy; reference
value)
±0.5% rdg. ±1 dgt.
Within ±6.0 V for 400 A/m AC (50 Hz/60 Hz) field
Add ±4.0 V to measured voltage values.
(Value describes effect when an adjacent wire
with a potential difference of 400 V AC is in
contact with the voltage sensor's clip.)
Add ±4.0 V to measured voltage values.
(When measuring insulated wires at a humidity
level of 70% RH to 80% RH)
Equivalent to ±2.6°
Normal (positive) phase, reversed (negative)
phase (3-phase/3-wire, 3-phase/4-wire)
1
2
3
4
5
6
7
47