Input Specifications and Measurement Specifications

Options
5.2 Input Specifications and
Measurement Specifications
Basic specifications
Measurement
parameters
Measurement targets
Measurable conductor
diameter
Voltage detection
method
Voltage sensor rated
voltage
Voltage measurement
method
Measurement display
Response time
Display update rate
Maximum rated input to
ground voltage
46
See p.  2 .
3-phase AC voltage (line voltage, voltage
relative to ground), frequency
Insulated wires*, metal parts*
*Not compatible with shielded wires.
3-phase 90.0 V AC to 520.0 V AC (45 Hz to 66
Hz)
Outer diameter: φ 6 mm to 30 mm
Coupled capacitance cancellation method
400 V AC relative to ground per voltage sensor
circuit
Digital sampling, true RMS method
4-digit 7-segment LCD × 3
3 sec. or less
500 ±10 ms
600 V AC (CAT IV)
Anticipated transient overvoltage: 8000 V