Aim-TTI PowerFlex QPX1200 Series Instruction Manual page 32

Table of Contents

Advertisement

V1O?
I1O?
DELTAV1 <nrf>
DELTAI1 <nrf>
DELTAV1?
DELTAI1?
INCV1
INCV1V
DECV1
DECV1V
INCI1
DECI1
OP1 <nrf>
OP1?
OPALL <nrf>
SENSE1 <nrf>
SAV1 <nrf>
RCL1 <nrf>
System and Status Commands
*CLS
*ESE <nrf>
*ESE?
*ESR?
*IST?
*OPC
*OPC?
Return the output readback voltage – response is <nr2>V<rmt> where
<nr2> is in Volts.
Return the output readback current – response is <nr2>A<rmt> where <nr2>
is in Amps.
Set the output voltage step size to <nrf> Volts.
Set the output current step size to <nrf> Amps.
Return the output voltage step size – response is DELTAV1 <nr2><rmt>
where <nr2> is in Volts.
Return the output current step size – response is DELTAI1 <nr2><rmt>
where <nr2> is in Amps.
Increment the output voltage by step size.
Increment the output voltage by step size with verify.
Decrement the output voltage by step size.
Decrement output voltage by step size with verify.
Increment the output current limit by step size.
Decrement the output current limit by step size.
Set output on/off where <nrf> has the following meaning: 0=OFF, 1=ON.
Returns output on/off status.
The response is <nr1><rmt> where 1 = ON, 0 = OFF.
Set all outputs on/off where <nrf> has the following meaning: 0=OFF, 1=ON.
In this instrument it has the same action as OP1 <nrf>.
Set output sense mode where <nrf> has the following meaning: 0=local,
1=remote.
Save the current PSU set up to the set-up store specified by <nrf>
where <nrf> can be 0-9.
Recall a set up for the PSU from the set-up store specified by <nrf>
where <nrf> can be 0-9.
Clear Status. Clears the Status structure: this indirectly clears the status byte
register.
Set Event Status Enable Register to the value <nrf>
Report value of Event Status Enable Register
Query and clear the Event Status Register. The response format is
<nr1><rmt>. See Status Reporting section for details of the response.
Returns ist local message as defined by IEEE Std. 488.2. The syntax of the
response is 0<rmt>, if the local message is false, or 1<rmt>, if the local
message is true.
Sets the Operation Complete bit (bit 0) in the Standard Event Status
Register. This will happen immediately the command is executed because of
the sequential nature of all operations.
Query Operation Complete status. The response is always 1<rmt> and will
be available immediately the command is executed because all commands
are sequential.
31