As required by the LXI standard, the instrument provides an XML identification document that can
be queried via a GET at "http://<hostname>:80/lxi/identification" that conforms to the LXI XSD
Schema (available at http://www.lxistandard.org/InstrumentIdentification/1.0) and the W3C XML
Schema Standards ( http://www.w3.org/XML/Schema ). This document describes the instrument.
TCP Sockets
The instrument uses 2 sockets on TCP port 9221 for instrument control and monitoring. Text
commands are sent to this port as defined in 'Remote Commands' and any replies are returned
via the same port. Any string must be one or more complete commands. Commands may be
separated with either semicolons ";" or line feeds. No terminator is required since the TCP frame
contains complete commands though commands may be sent with a terminator if desired (it will
be ignored). Each command over TCP behaves as if it is terminated with a command terminator
(ASCII character 0AH, line feed).
GPIB Interface
The GPIB interface 24-way connector is located on the instrument rear panel. The pin
connections are as specified in IEEE Std. 488.1-1987 and the instrument complies with IEEE Std.
488.1-1987 and IEEE Std. 488.2-1987.
GPIB Subsets
This instrument contains the following IEEE 488.1 subsets:
Source Handshake
Acceptor Handshake
Talker
Listener
Service Request
Remote Local
Parallel Poll
Device Clear
Device Trigger
Controller
Electrical Interface
Query Error Register - GPIB IEEE Std. 488.2 Error Handling
The IEEE 488.2
If the instrument is addressed to talk and the response formatter is inactive and the input queue is
empty then the
the Standard Event Status Register, a value of 3 to be placed in the Query Error Register and the
parser to be reset. See the Status Reporting section for further information.
The IEEE 488.2
send a response message and a
or the input queue contains more than one END message then the instrument has been
INTERRUPTED
Standard Event Status Register, a value of 1 to be placed in the Query Error Register and the
response formatter to be reset thus clearing the output queue. The parser will then start parsing
the next
<PROGRAM MESSAGE UNIT>
further information.
24
error (addressed to talk with nothing to say) is handled as follows.
UNTERMINATED
error is generated. This will cause the Query Error bit to be set in
UNTERMINATED
error is handled as follows. If the response formatter is waiting to
INTERRUPTED
<PROGRAM MESSAGE TERMINATOR>
and an error is generated. This will cause the Query Error bit to be set in the
SH1
AH1
T6
L4
SR1
RL2
PP1
DC1
DT0
C0
E2
from the input queue. See the Status Reporting section for
has been read by the parser