Thermo Scientific 700 Series Operating And Maintenance Manual page 39

Forma -86c ult freezer
Hide thumbs Also See for 700 Series:
Table of Contents

Advertisement

Section 6
Specifications
6-2
Model 700 Series
Model
705
Temperature
-50C to -86C (-58F to -123F) in an 18C to 32C* (64.4F to 89.6F) ambient
Range
33.3"W x 77.8"H x 37.0"
Exterior
Dimensions
84.6 x 197.6 x 94.0cm
23.0"W x 51.5"H x 25.3"
Interior
Dimensions
58.4 x 130.8 x 64.3cm
17.3 cu. ft.
Capacity
(489.9 liters)
Refrigeration
Two 1 HP (2545 BTUH each)
Insulation
Non-CFC, foamed-in-place urethane: 5.0" (12.7 cm) cabinet; 4.5" (11.4 cm) door
230V,50/60Hz, 12.0FLA
Electrical
Operating Range:
208VAC-240VAC
15 amp, 230V,
Dedicated Circuit,
Breaker
Requirements
15 Amp
Time Delay Breaker
Shipping Weight:
795 lbs. (360.6 kg)
Motor
*Compressors may not cycle off with cabinet running at -86C in a 32C ambient.
706
756
40.8"W x 77.8"H x 37.0"
40.8"W x 77.8"H x 37.0"
103.6 x 197.6 x 94.0cm
103.6 x 197.6 x 94.0cm
30.6"W x 51.5"H x 25.3"
30.6"W x 51.5"H x 25.3"
77.7 x 130.8 x 64.3cm
77.7 x 130.8 x 64.3cm
23.0 cu. ft.
23.0 cu. ft.
(651.3 liters)
(651.3 liters)
230V,50/60Hz, 12.0FLA
120V, 60 Hz, 16.0 FLA
Operating Range:
Operating Range:
208VAC-240VAC
108-130V
15 amp, 230V,
20 amp, 120V,
Dedicated Circuit,
Dedicated Circuit,
15 Amp
20 amp
Time Delay Breaker
Time Delay Breaker
900 lbs. (408.2 kg)
900 lbs. (408.2 kg)
Thermo Fisher Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

702703704705706756