Thermo Scientific HIGHPlate™ Instruction Manual 50119962-e • 05 / 2020 Visit us online to register your warranty: thermofisher.com/labwarranty...
Page 2
WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:...
Page 3
Issue Date: January 2009 Test Summary A Thermo Scientific 75003606 contained rotor (Max speed 6,300 rpm) was supplied by Thermo Fisher and containment tested at 6,300 rpm using the method described i n Annex AA of EN 61010-2-020. The rotor was shown to contain a spill when tested i n triplicate.
Page 5
Rotor Loading ..........................4-1 Before a Run ......................4-2 Proper Loading ......................4-2 Improper Loading ....................4-2 Maximum Loading ....................4-3 Cycle Counter ......................4-3 Chapter 5 Aerosol-tight Applications ....................5-1 Basic Principles ......................5-2 Placing O-ring ......................5-2 Fill Level ........................5-2 Checking the Aerosol-Tightness ................5-2 ™ Thermo Scientific HIGHPlate...
Page 6
Chapter 6 Maintenance and Care ......................6-1 Cleaning intervals ....................6-2 Cleaning ........................6-2 Disinfection ......................6-3 Decontamination ..................... 6-4 Autoclaving ......................6-5 Service of Thermo Fisher Scientific ................6-5 RCF-values ........................A-1 Chemical Compatibility Chart ................B-1 ™ Thermo Scientific HIGHPlate...
Page 7
• Do not remove the magnet at the rotor bottom. • Do not use rotors which show any signs of corrosion and / or cracks. • Use only with rotors which have been loaded properly. • Never overload the rotor. ™ Thermo Scientific HIGHPlate...
Page 8
CAUTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. ™ Thermo Scientific HIGHPlate...
Page 10
6168 Max. / min. radius [cm] 13.9 / 5.8 13.9 / 5.8 Accel. / braking time [s] 60 / 65 60 / 65 Aerosol-tight Permissible temperature range autoclavable (cycle number) °C Tested by HPA, Porton-down, UK ™ Thermo Scientific HIGHPlate...
Page 11
6168 Max. / min. radius [cm] 13.9 / 5.8 13.9 / 5.8 Accel. / braking time [s] 90 / 65 90 / 65 Aerosol-tight Permissible temperature range autoclavable (cycle number) °C Tested by HPA, Porton-down, UK ™ Thermo Scientific HIGHPlate...
Page 12
6168 Max. / min. radius [cm] 13.9 / 5.8 13.9 / 5.8 Accel. / braking time [s] 60 / 65 60 / 65 Aerosol-tight Permissible temperature range autoclavable (cycle number) °C Tested by HPA, Porton-down, UK ™ Thermo Scientific HIGHPlate...
Page 13
6168 Max. / min. radius [cm] 13.9 / 5.8 13.9 / 5.8 Accel. / braking time [s] 90 / 65 90 / 65 Aerosol-tight Permissible temperature range autoclavable (cycle number) °C Tested by HPA, Porton-down, UK ™ Thermo Scientific HIGHPlate...
Page 14
6168 Max. / min. radius [cm] 13.9 / 5.8 13.9 / 5.8 Accel. / braking time [s] 90 / 65 90 / 65 Aerosol-tight Permissible temperature range autoclavable (cycle number) °C Tested by HPA, Porton-down, UK ™ Thermo Scientific HIGHPlate...
Page 15
6300 Maximum RCF value at n 6168 Max. / min. radius [cm] 13.9 / 5.8 Accel. / braking time [s] 60 / 65 Aerosol-tight Permissible temperature range autoclavable (cycle number) °C Tested by HPA, Porton-down, UK ™ Thermo Scientific HIGHPlate...
Page 18
2. Make sure you use a rubber bottom placed in the cut outs of the micro plate carrier. Deepwell plates may be used without the rubber bottom. 3. Make sure the load is balanced. 4. Place the loaded micro plate carriers in the rotor. ™ Thermo Scientific HIGHPlate...
Page 20
WARNING If the rotor cannot be properly locked in place after several attempts, then the ™ AutoLock is defective and you are not permitted to operate the rotor. Check for any damage to the rotor, damaged rotors must not be used. Keep the hub area clear of objects. ™ Thermo Scientific HIGHPlate...