BPBIO320S User Manual for Measurement Guide and Setup
Thank you for purchasing the BPBIO320S. This user's manual describes all the features of the
BPBIO320S. Please read before use and keep it in a safe place. By following the manual instructions,
you will be able to use the BPBIO320S more safely and effectively.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BPBIO320S and is the answer not in the manual?

Questions and answers

Summary of Contents for inbody BPBIO320S

  • Page 1 Thank you for purchasing the BPBIO320S. This user’s manual describes all the features of the BPBIO320S. Please read before use and keep it in a safe place. By following the manual instructions, you will be able to use the BPBIO320S more safely and effectively.
  • Page 2 Le non-respect des consignes de sécurité peut endommager l'équipement. Remarque Le but est d'aider les utilisateurs à comprendre. For problems with the BPBIO320S or any clinical questions, contact: Representative & Sponsor Information InBody Europe B.V. Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands...
  • Page 3 Intended Use The BPBIO320S is a digital monitor intended for use in measuring blood pressure and pulse rate in user population with upper arm circumference ranging from 17cm to 42cm (7-inch to 17-inch). The systolic blood pressure and diastolic blood pressure are measured by non-invasive blood pressure (“NIBP”) measuring method.
  • Page 5: Table Of Contents

    User’s Manual Contents I. BPBIO320S Installation A. Product Components B. Operating Environment C. Installation Instructions D. Loading and Reloading Printer Paper E. Replacing Cuff Cover F. Setup G. Connecting Multi-Display Device H. Connecting to InBody I. Precautions for Maintenance II. Blood Pressure Test A.
  • Page 6: Bpbio320S Installation

    Plug Type B or F B. Operating Environment Please make sure the environment is adequate for the BPBIO320S installation. This equipment is designed for indoor use. If installing the equipment outdoors, the following requirements must be fulfi lled: Temperature range 5 to 40ºC (41 to 104ºF)
  • Page 7: Installation Instructions

    C. Installation Instructions 1. Open the BPBIO320S box, and remove the air cushion (small size). Unfold the air cushion (large size) and take out the BPBIO320S from the box, and then remove the protective box. Take off the protective vinyl out of the BPBIO320S, and remove the packing material.
  • Page 8 2. Place the equipment on a solid table that can withstand the weight of the BPBIO320S (approx. 9kg (20 lb)). If desk for the BPBIO320S is available, align the grooves on the bottom of the BPBIO320S with the grooves on the desk and secure them with two fi xing screws.
  • Page 9 Do not disassemble the equipment arbitrarily. This may cause electric shock, injury, product malfunction, and/ • or inaccurate results. Unauthorized disassembly will void the manufacturer’s warranty. Do not let the BPBIO320S touch with other electronic devices when on. This may cause electric shock. • Avertissement N'installez pas l'équipement à...
  • Page 10 6. Turn the BPBIO320S on by pressing the power switch located on the back of the device. Power Switch 7. When the BPBIO320S is turned on, the following screen is displayed and a calibration is automatically performed. Caution Do not operate or touch the device while it is calibrating. It may cause errors to future tests.
  • Page 11: Loading And Reloading Printer Paper

    Paper cut with scissors Ripped paper 2) Open the cover by pressing ‘PUSH’ while the BPBIO320S is turned on. 3) If the device is being used for the first time, remove the protective paper located inside. * The protective paper (thermal printing paper) is in the printer groove. Pull upward to remove it.
  • Page 12 6) Remove the cut paper, and close the printer paper cover. 2. Reloading Printer Paper 1) Open the cover by pressing ‘PUSH’ mark on the bottom center of it while the BPBIO320S is turned 2) Press button to remove the remaining paper.
  • Page 13: Replacing Cuff Cover

    The cuff cover of the BPBIO320S should be replaced periodically for sanitary purposes. Follow the guide below to replace the cuff cover. 1. Remove the front cover of the BPBIO320S by pushing downward. Loosen the screw with a screwdriver. Screws Front Cover 2.
  • Page 14 5. Secure the new cuff cover onto the grooves on the front and rear of the BPBIO320S. 6. Attach the previously removed front and rear rings on the BPBIO320S and secure them with screws. Rear ring Front ring 7. Close the front cover of the BPBIO320S.
  • Page 15: Setup

    Note Contents of results sheet output are only applicable to BPBIO320S. • Remarque • Le contenu de la sortie de la feuille de résultats ne s’applique qu’à BPBIO320S. 1. Settings 1) When button and button are pressed and held at the same time for 2 seconds while on the home screen, the fi rst screen of the Setup, 'Sound Setting' is displayed.
  • Page 16 The following is an example showing how to change a ‘manual output’ Automatic printing options. While on the home screen, press and hold button and button at the same time for 2 seconds. ▼ Press button to move to screen. Press button to select ▼...
  • Page 17 2. Detailed description of Setup When in the Setup menu, the following information is displayed in the Systolic, Diastolic and Pulse sections of the screen. Display Function Description Systolic Diastolic Pulse Reading test instruction and measurement results Reading test instruction only Sound Reading measurement results only Beep guide...
  • Page 18 Output of caution comment Caution commemt Not used Output of pulse graph PULSE graph Not used Result Mean arterial pressure/PULSE pressure sheet item /Pressure rate product output Additonal result values Not used Output of QR code* QR code Not used 3 consecutive measurements mode Measurement 3 consecutive...
  • Page 19 3. Blood Pressure Evaluation 1) While on the home screen, press and hold button and UP (▲) button at the same time for more than 2 seconds. 2) When 111 is displayed in the systolic blood pressure column, diastolic blood pressure column, and pulse column as shown below, the blood pressure evaluation has been disabled.
  • Page 20 4. Displaying test counts 1) While on the home screen, hold the DOWN (▼) button for more than 2 seconds. Test counts for the day will be displayed in the diastolic blood pressure column. Test counts for the day and total test counts are printed.
  • Page 21: Connecting Multi-Display Device

    2. Connect the multi-display device cable to the SUB DISPLAY connection terminal on the back of the device while the BPBIO320S is turned off . Then connect the other end of the display device cable to connection terminal on the left of the display device.
  • Page 22: Connecting To Inbody

    1. Turn off the BPBIO320S. * If the BPBIO320S is already turned on while being connected to the InBody, the connection may not be smooth. 2. Connect the serial cable to the RS232C (9-pin serial) connection terminal located on the back of the BPBIO320S.
  • Page 23: Precautions For Maintenance

    Make sure the power cord is not a tripping hazard. • The BPBIO320S calibrates automatically when it is turned on. But it may be necessary for a specialist to inspect the device once every two years to maintain its performance and safety. If an error message appears during the selftest, please contact product support.
  • Page 24: Blood Pressure Test

    • Évitez de parler ou de bouger durant le test. • BPBIO320S est un périphérique de test dédié au bras droit. Les résultats du test peuvent être inexacts si la • pression artérielle est examinée avec le bras gauche. B. Test Instructions 1.
  • Page 25 • of the BPBIO320S. The cuff will release quickly. If the air is not released after the button is pressed, turn off the BPBIO320S. • Attention En cas d'urgence pendant le test, appuyez sur le bouton ou sur le bouton en bas à...
  • Page 26 6. The cuff is automatically released after completing the test. Wait for the cuff to fully loosen and pull out your arm. 7. Check the results on the screen. * The voice guide will be heard and the Result Sheet will be printed depeneding on settings. * Do not pull the Result Sheet while it is printing.
  • Page 27 * The user can check results by using 'Check Blood Pressure Result' at the bottom right of the Information Panel * Consult a physician for accurate diagnosis. Check Blood Pressure Check Blood Pressure Result Result Note Rest for 5 minutes before testing again. •...
  • Page 28: Test Posture

    Sit as as close as possible to the BPBIO320S * Be sure to place the elbow in the elbow groove before starting the test. * Remove thick clothes. Do not roll up sleeves.
  • Page 29: Transportation And Storage

    III. Transportation and Storage A. Cautions during Transportation To transport the BPBIO320S safely, two people should keep the equipment parallel to the ground. Caution When transporting the equipment alone, keep the equipment horizontal while transporting. • Attention Lorsque vous transportez l'équipement seul de façon inévitable, soulevez-le en le maintenant horizontal comme suit.
  • Page 30: Repacking Instructions

    Protective box Package protective vinyl 3. Place the large air cushion on the bottom of the packing box and place the BPBIO320S on it. Cover with the small air cushion and seal the box with tape. Air cushion (large size)
  • Page 31: Faq

    The error messages can be displayed on the screen if they occur during the test. Please provide the error code to Product Support. B. Regarding the Equipment If problems cannot be resolved with the information below, please contact InBody. Problem Cause Corrective measures Plug the power cable into a grounded 3-prong outlet.
  • Page 32 • Éteignez complètement l’appareil lorsque vous devez remplacer le fusible. • Veillez à utiliser un fusible de calibre approprié (250V, 2.5 A) Turn on the volume knob on the rear of the BPBIO320S. The volume knob on the rear of the blood pressure monitor may be turned off .
  • Page 33 The company is not liable for the problems occurred when others repair and inspect the equipment except • employees of InBody Co., Ltd. Attention InBody Co, Ltd se décharge de toute responsabilité en cas de problème si la maintenance et l’inspection de • votre appareil ne sont pas. Note Since error messages and things of detail such as incorrect result sheets can be used as a basis for A/S, please •...
  • Page 34: Regarding The Test

    The user may have tested without sufficient rest. Take the test after resting at • least 5 minutes. Do not talk or move during the test. • * Blood pressure may vary throught the day. InBody recommends testing at diff erent times of the day.
  • Page 35 If you had blood drawn or had an injection given in the either arm, it is recommended to wait at least 1 hour before testing, or measure with another arm. Can results vary depending BPBIO320S can operate at 5 to 40ºC (41 to 104ºF). However, if • • on the testing environment...
  • Page 36: Others

    Others * The BPBIO320S was manufactured by the quality control procedures of InBody Co., Ltd. InBody Co., Ltd. complies with ISO9001 and ISO13485 which are international quality management systems. * This equipment complies with IEC60601-1, the international safety standard for electronic medical equipment. This equipment also complies with IEC60601-1-2, the international standard for electromagnetic compatibility.
  • Page 37: Safety Information

    B. Safety Information Warning If this equipment has been altered, it must be inspected to ensure that it can be used safely. • Results can only be interpreted by experienced healthcare professionals. It cannot be used for diagnosis, • medication, or other treatments performed at the consumer’s discretion without a doctor ̓ s prescription. Avertissement Si cet équipement a été...
  • Page 38: Product Classification

    Other Symbols Manufacturer Serial number Authorized representative in the Alternating current EUROPEAN COMMUNTY European conformity Operating instructions Complies with AAMI ES60601-1 CSA-C22.2 No. 60601-1 (MET mark only applies for USA and Canada.) C. Product Classifi cation Oscillometric multi-functional electronic blood pressure monitor Type of protection against electric shock: Class 1 device •...
  • Page 39: Specifications

    D. Specifications Test Instructions Oscillometric Test range Pressure: 0-300 mmHg Pulse: 30-240 bpm Degree of Precision Pressure: ±2 mmHg Pulse: ±1.5% Test result Systolic blood pressure, diastolic blood pressure, Mean arterial pressure, Pulse pressure, PULSE rate, RPP (rate pressure product) Test Time Approx.
  • Page 40: Emc Declaration

    E. EMC Declaration Test Name IEC 60601- Ref. Standard Ports to Test AC Mains Test Level Required Notes Voltage 4th Edition 100V, 50Hz Mains terminal CISPR 100V, 60Hz Group 1, AC Mains disturbance voltage 11:2015/ 120V, 60Hz Class B AMD2:2019 220V, 60Hz 230V, 50Hz 100V, 50Hz...
  • Page 41 100V, 50Hz 100V, 60Hz Power-frequancy IEC 61000-4- Enclosure 120V, 60Hz 30 A/m Magnetic Field 8:2009 220V, 60Hz 50 Hz & 60 Hz 230V, 50Hz Battery AT 0°, 45°, 0% UT for 0.5 cycle 90°, 135°, 180°, 225°, 270°, 315° 0% UT for 1 cycle AT 0°, 180°...
  • Page 42 Electromagnetic immunity The [Model name] is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the [Model name]. Otherwise, degradation of the performance of this equipment could result.
  • Page 46 1639 ©2022 InBody Co., Ltd. All rights reserved. IM-ENG-E5-J-220629...