The device is not
turning on.
Voice guide does not
work.
32
Cause
Turn off the BPBIO320S, and replace the fuse.
The fuse socket is located on the back of the device beneath
the power plug.
Pull the fuse out by using a flathead screwdriver. Replace
the fuse with the provided spare fuse. Additonal fuses can
be purchased if necessary.
The fuse is blown.
• Be sure to turn off the power when replacing the fuse.
• The correct rated fuse must be used. (250V, 2.5AL)
• Éteignez complètement l'appareil lorsque vous devez
remplacer le fusible.
• Veillez à utiliser un fusible de calibre approprié (250V, 2.5 A)
Turn on the volume knob on the rear of the BPBIO320S.
The volume knob on
the rear of the blood
pressure monitor
may be turned off .
If the sound setting
The sound setting
(alarm sound) will be heard. If it is set at
has been set to 'Beep
will be heard. If you want the voice guide to be heard, set the
guide' or 'Disable
sound setting to
speaker.'
* Refer to Setup on page 13 to change the sound settings.
Measures
Spare fuse
Warning
Avertissement
is set at
.
TYPE: Fast-Acting
Rated current: 2.5AL
Rated voltage: 250V
, only the 'beep' sound
, no sound