2. Place the equipment on a solid table that can withstand the weight of the BPBIO320S (approx. 9kg (20 lb)).
If desk for the BPBIO320S is available, align the grooves on the bottom of the BPBIO320S
with the grooves on the desk and secure them with two fi xing screws.
* Desk for the BPBIO320S sold separately.
Caution
Install the BPBIO320S on a flat, vibration-free floor.
If the equipment is installed where the floor is not flat, it may topple during a test or the test results may be
inaccurate.
Attention
Installez BPBIO320S sur un sol plat et sans vibrations.
Si l'équipement est installé sur un sol qui n'est pas plat, il peut basculer pendant un test ou les résultats du test
peuvent être inexacts.
3. Mount the Information Panel onto the BPBIO320S.
4. Connect the power cable provided with the BPBIO320S. Plug the power cable into a grounded 3-prong outlet.
* The anti-theft groove can be used to secure the equipment to the table.
<Plug Type B>
3-terminal outlet Type B
8
Desk fi xing grooves
Anti-theft groove
Grounded
terminal
Desk grooves
<Plug Type F>
3-terminal outlet Type F
Fixing screws
Grounded terminal