Siemens SITRANS L Operating Instructions Manual page 137

Radar transmitters with foundation fieldbus
Hide thumbs Also See for SITRANS L:
Table of Contents

Advertisement

Before Auto False Echo Suppression
① TVT Hover Level
② default TVT
③ false echo
1. Determine Auto False Echo Suppression Range. Measure the actual distance from the sensor
reference point to the material surface using a rope or tape measure.
2. Subtract 0.5 m (20") from this distance, and use the resulting value.
After Auto False Echo Suppression
SITRANS LR250 with FOUNDATION FIELDBUS
Operating Instructions, 09/2021, A5E32221411-AH
echo marker
material level
false echo
learned TVT
Auto False Echo Suppression Range
material level
default TVT
echo marker
Parameter reference
137