Calibration Specifications - MSA ALTAIR 5X PID Operating Manual

Multigas detector
Hide thumbs Also See for ALTAIR 5X PID:
Table of Contents

Advertisement

US
7.3

Calibration Specifications

Sensor
*ClO
2
Cl
2
CO
COMB Butane (1.4 % Vol)
COMB Hydrogen (4,0 % Vol) Fresh Air
COMB Methane (0 - 5 % Vol) Fresh Air
COMB Methane (4.4 % Vol)
COMB Methane (5 % Vol)
COMB Pentane
COMB Propane (1.7 % Vol)
COMB Propane (2.1 % Vol)
H
S
2
HCN
NH
3
NO
NO
2
O
2
PH
3
PID
SO
2
Span values can be changed if using different gas cylinders than those listed. Changes can be made
using MSA Link software and through calibration cylinder setup.
*For most accurate results, calibration with ClO
**Zero cal time is one minute if a catalytic combustible sensor is installed - 30 seconds if not.
LEL values, if not listed here, are according to EN 60079-20-1. Local regulations may differ.
65
Zero Gas
Zero Cal
Value**
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
0
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
20.8 %
Fresh Air
0
Fresh Air
0
Fresh Air
0
is recommended.
2
ALTAIR 5X PID
Span Cal Gas
Span Cal
Value
2 ppm Cl
0.8 ppm
2
10 ppm Cl
10 ppm
2
60 ppm CO
60 ppm
1.45 % Vol Methane 46 % LEL
1.45 % Vol Methane 33 % LEL
2.5 % Vol Methane
2,5 %
1.45 % Vol Methane 33 % LEL
1.45 % Vol Methane 29 % LEL
1.45 % Vol Methane 58 % LEL
1.45 % Vol Methane 37 % LEL
1.45 % Vol Methane 46 % LEL
20 ppm H
S
20 ppm
2
10 ppm HCN
10 ppm
25 ppm NH
25 ppm
3
50 ppm NO
50 ppm
10 ppm NO
10 ppm
2
15 % O
15 %
2
0.5 ppm PH
0.5 ppm
3
100 ppm
100 ppm
isobutylene
10 ppm SO
10 ppm
2
7 Technical Specifications
Time
(min)
6
2
1
1
1
1
1
1
1
1
1
1
4
2
4
2
1
1
1
1

Advertisement

Table of Contents
loading

Table of Contents