Calibration Check; Measuring Gas Concentrations; Combustible Gases (% Lel) (Figure 2-2) - MSA Solaris Operating Manual

Multigas detector
Hide thumbs Also See for Solaris:
Table of Contents

Advertisement

Available languages

Available languages

Calibration Check

The calibration check is simple and should only take about one minute.
Perform this calibration check before each day's use.
1. Turn ON the Solaris Multigas Detector in clean, fresh air.
2. Verify that readings indicate no gas is present.
3. Attach calibration cap to the Solaris Multigas Detector.
4. Ensure that
TOP
"
that "TOP" is positioned at the top of the instrument.
5. Attach regulator (supplied with calibration kit) to the cylinder.
6. Connect tubing (supplied with calibration kit) to the regulator.
7. Attach other end of tubing to the calibration cap.
8. Open the valve on the regulator.
The regulator flow rate is 0.25 lpm.
The reading on the Solaris Multigas Detector display should be
within the limits stated on the calibration cylinder or limits
determined by your company.
If necessary, change cylinder to introduce other calibration
gases.
If readings are not within these limits, the Solaris Multigas
Detector requires recalibration. See Chapter 4, "Calibration."

Measuring Gas Concentrations

Combustible Gases (% LEL) (FIGURE 2-2)

Figure 2-2. Instrument in LEL Alarm
2-6
and
on the calibration cap are oriented so
"
"
"

Advertisement

Table of Contents
loading

Table of Contents