Never let the end of the sampling line touch or go under any liquid surface. If liquid is drawn into the device, readings will be
inaccurate and device could be damaged. We recommend the use of an MSA sample probe containing a special membrane
filter, permeable to gas but impermeable to water, to prevent such an occurrence.
During operation, a pump alarm may occur when the:
• Flow system is blocked
• Pump is inoperative
• Sample lines are attached or removed.
To Clear Pump Alarm
1. Correct any flow blockage.
2. Press the p button.
The Pump will now restart.
6.3

Replacing the Battery

WARNING!
Never replace the battery in a hazardous area. This could result in an explosion.
Failure to follow this warning can result in serious personal injury or death.
Figure 17 Battery Replacement
1
Captive screw
2
1. Unscrew the two captive screws on the rear of the device.
2. Pull the battery pack out of the device by gripping the sides and lifting it up and away from the device.
3. When replacing the battery, be sure to accurately align screws and battery with housing.
4. Screws should be tightened and torqued to 5.5 in lb.
6.4

Maintenance Procedure - Replacing or Adding a Sensor

Any factory-installed Series 20 sensor may be removed or replaced with another Series 20 sensor of the same gas type. Any
XCell sensor may be removed or replaced according to the positions allowed in table after
If the type of any sensor is to be changed, the device must be returned to an authorized service center.
51
Battery pack
ALTAIR 5X PID
6 Maintenance
Figure 18
.