Installing/Replacing the Battery

Remove belt clip by pressing the two tabs near
the top and releasing from the back. Remove
the two slotted head screws on the back of the
case, located under the clip. Remove the 9-V
battery from the snap connector and replace
with a fresh 9-V battery or install a fresh 9-V
battery for first time use. Replace the battery
cover and the two slotted screws and snap the
belt clip back. Wait one minute before turning
the instrument on and rezeroing it. The
instrument will not turn ON for 45 seconds. The
backlight is ON and --- is displayed during this
period.

" CAUTION

When installing the battery, be careful not
to press the Zero button. Pressing Zero
within one minute of battery installation can
cause an incorrect zeroing; the instrument
will show a positive value in fresh air and
must to be re-zeroed.

Replacing Sensor Assembly

Remove phillips screw from sensor assembly.
Rotate approximately 15
assembly out. Remove the shorting bar from the
new assembly. Place the assembly into the
instrument. Rotate approximately 15
counterclockwise. Insert phillips screw and
tighten. The sensor assembly can be opened if
you choose to replace only the sensor.

Hot Gas Sampling

(for use with MiniCO Responders only)
1. Connect Hot Gas Sampler (PN 803848) to the
sampling line.
2. Connect Aspirator Assembly (P/N 809964) to
the sampling line.
3. Attach the aspirator to Calibration Cap
(P/N 710492).