Calibration Adjustment

Calibration adjustment is done electronically by internal software with the Product case
closed. The recommended adjustment procedure is a 2-point adjust that adjusts a low and
high point. This adjustment can be applied to any of the 700 family of gauges. It is the
only adjustment that may used with a 700RG ranges.
This is an ambient temperature calibration, and must be done at an ambient
temperature of 23 °C ± 3 °C (72 °F ± 5 °F). Calibration out of this
temperature range voids the temperature compensation software in the
Product.

Test Equipment

For verification and calibration adjustment, pressure and/or vacuum standards that can
make and show pressures from vacuum to the full-scale range of the UUT are necessary.
A TUR standard of 4:1 or better is necessary to keep the Product at its specified accuracy.

Connections

The Product uses a ¼ inch NPT male connection in the pressure input port. Different
adapters can be necessary to connect to the pressure standard. Always make sure the
hose, tubing, and fittings. Have a rated working pressure at or above the pressure of the
unit. It is also important that there be no leaks when you calibrate the Product. Use Teflon
tape where necessary.

Enter 2-Point Adjust Mode

Follow the instructions for serial interface setup in the next section. Figure 2 can be used
to guide setup if HyperTerminal will be used.